Reference Manual 00809-0100-4308, Rev BA December 2014 Rosemount 3308 Series Wireless Guided Wave Radar, 3308A Reference Manual
Reference Manual Title Page 00809-0100-4308, Rev BA December 2014 Rosemount 3308 Series Wireless Guided Wave Radar Level Transmitter NOTICE Read this manual before working with the product. For personal and system safety, and for optimum product performance, make sure you thoroughly understand the contents before installing, using, or maintaining this product. The United States has two toll-free assistance numbers and one international number. Customer Central 1-800-999-9307 (7:00 a.m. to 7:00 p.m.
Title Page Reference Manual December 2014 00809-0100-4308, Rev BA Electrical shock can result in death or serious injury. Ground device on non-metallic tanks (e.g. fiberglass tanks) to prevent electrostatic charge build-up. Single lead probes are sensitive for strong electromagnetic fields and therefore not suitable for non-metallic tanks. Care must be taken during transportation of power module to prevent electrostatic charge build-up.
Reference Manual Title Page 00809-0100-4308, Rev BA December 2014 This device complies with Part 15 of the FCC Rules. Operation is subject to the following conditions: This device may not cause harmful interference. This device must accept any interference received, including interference that may cause undesired operation. This device must be installed to ensure a minimum antenna separation distance of 20 cm (8 in.) from all persons.
Title Page December 2014 vi Reference Manual 00809-0100-4308, Rev BA
Reference Manual Table of Contents 00809-0100-4308, Rev BA December 2014 Contents 1Section 1: Introduction 1.1 Service support. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.2 Product recycling/disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 2Section 2: Transmitter Overview 2.1 Theory of operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Reference Manual Table of Contents 00809-0100-4308, Rev BA December 2014 3.8 Install the power module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38 3.9 Position the antenna . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38 3.10Utilize the device display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39 3.10.1Rotate the device display . . . . . . . . . . .
Reference Manual Table of Contents 00809-0100-4308, Rev BA December 2014 6Section 6: Service and Troubleshooting 6.1 Safety messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69 6.2 Alert messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72 6.2.1 Device display alerts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72 6.2.
Reference Manual Table of Contents 00809-0100-4308, Rev BA December 2014 A.2.2 Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107 A.2.3 Accuracy over measuring range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108 A.2.4 Maximum measuring range and minimum dielectric constant . . . . . . . 110 A.2.5 Interface measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110 A.2.
Reference Manual Table of Contents 00809-0100-4308, Rev BA December 2014 C.2.5 Mounting bracket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152 C.2.6 Antenna . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152 C.3 Review installation considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153 C.3.1 Antenna mounting . . . . . . . . . . . . . . . . . . . . . . . . .
Table of Contents December 2014 vi Reference Manual 00809-0100-4308, Rev BA Table of Contents
Reference Manual Section 1: Introduction 00809-0100-4308, Rev BA December 2014 Section 1 1.1 Introduction Service support To expedite the return process outside of the United States, contact the nearest Emerson Process Management representative. Within the United States, call the Emerson Process Management Instrument and Valves Response Center using the 1-800-654-RSMT (7768) toll-free number. This center, available 24 hours a day, will assist you with any needed information or materials.
Section 1: Introduction December 2014 2 Reference Manual 00809-0100-4308, Rev BA Introduction
Reference Manual Section 2: Transmitter Overview 00809-0100-4308, Rev BA December 2014 Section 2 Transmitter Overview Theory of operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Emerson's Smart Wireless . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Application characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Section 2: Transmitter Overview Reference Manual December 2014 00809-0100-4308, Rev BA Reference peak This peak is caused by the transition between transmitter and the tank vapor space or air. It is used by the transmitter as a starting reference point for distance to the level surface. Product surface peak This peak indicates the product level and is caused by a reflection from the product surface. Interface peak This peak indicates the interface level.
Reference Manual Section 2: Transmitter Overview 00809-0100-4308, Rev BA December 2014 Tank height The Tank Height is the distance from the Upper Reference Point to the Zero Reference Point. The transmitter measures the distance to the product surface and subtracts this value from the Tank Height to determine the level. Probe length The probe length is the distance between the Upper Reference Point and the end of the probe. If a weight is used at the end of the probe it shall be included.
Section 2: Transmitter Overview December 2014 2.2 Reference Manual 00809-0100-4308, Rev BA Emerson's Smart Wireless Emerson Smart Wireless is a self-organizing network solution. Wireless field instruments send data to a Gateway, directly or routed through any of the wireless devices in the network, as illustrated in Figure 2-4. Multiple communication paths are managed and analyzed in parallel to assure optimal communication and sustained network reliability even if obstructions are introduced.
Reference Manual Section 2: Transmitter Overview 00809-0100-4308, Rev BA 2.3.2 December 2014 In-tank obstructions The Rosemount 3308 Series Transmitter is relatively insensitive to objects in the tank since the radar signal is transmitted along a probe. Avoid physical contact between probes and agitators as well as applications with strong fluid movement unless the probe is anchored.
Section 2: Transmitter Overview Reference Manual December 2014 00809-0100-4308, Rev BA Emulsion layers Sometimes there is an emulsion layer (mix of the products) between the two products which, depending on its characteristics, will affect interface measurements. Please consult factory for guidelines on how to handle emulsion layers. 2.
Reference Manual Section 2: Transmitter Overview 00809-0100-4308, Rev BA 2.5 December 2014 Components of the transmitter The Rosemount 3308 Series transmitter housing contains advanced electronics for signal processing. The transmitter housing is made of aluminum or stainless steel, depending on specified option code. The radar electronics produces an electromagnetic pulse which is guided by the probe.
Section 2: Transmitter Overview Reference Manual December 2014 2.6 00809-0100-4308, Rev BA Probe selection guide for the Rosemount 3308 Series Transmitter Use the following guidelines to choose appropriate probe for your Rosemount 3308 Series transmitter: Table 2-1. Probe Selection Guide.
Reference Manual Section 2: Transmitter Overview 00809-0100-4308, Rev BA December 2014 Flexible Rigid single single lead lead Liquid or vapor spray might touch probe Flexible twin lead Coaxial NR NR NR G AD(3) G AD(3) G Turbulent conditions causing breaking forces AD NR AD NR Non-metallic tanks or open atmosphere applications AD(4) AD(4) AD(4) G High turbulence (1) Not in fully submerged applications.
Section 2: Transmitter Overview December 2014 12 Reference Manual 00809-0100-4308, Rev BA Transmitter Overview
Reference Manual Section 3: Installation 00809-0100-4308, Rev BA December 2014 Section 3 Installation Safety messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Installation procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Review mounting considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Section 3: Installation December 2014 Reference Manual 00809-0100-4308, Rev BA Explosions could result in death or serious injury. Installation of device in an explosive environment must be in accordance with appropriate local, national and international standards, codes, and practices. Ensure device is installed in accordance with intrinsically safe or non-incendive field practices. Electrical shock can result in death or serious injury. Ground device on non-metallic tanks (e.g.
Reference Manual 00809-0100-4308, Rev BA Section 3: Installation December 2014 To ensure long life for your radar transmitter, and to comply with hazardous location installation requirements, tighten covers on both sides of the electronics housing. Any substitution of non-recognized parts may jeopardize safety. Repair, e.g. substitution of components etc., may also jeopardize safety and is under no circumstances allowed. Use caution when handling the Power Module.
Section 3: Installation Reference Manual December 2014 3.
Reference Manual Section 3: Installation 00809-0100-4308, Rev BA 3.3 December 2014 Review mounting considerations Before installing the Rosemount 3308 Series Transmitter, consider recommendations for sufficient free space, mounting position and special needs for non-metallic tanks. 3.3.1 Recommended mounting position When finding an appropriate mounting position for the transmitter, the conditions of the tank must be carefully considered.
Section 3: Installation Reference Manual December 2014 00809-0100-4308, Rev BA Make sure the nozzle does not extend into the tank. Note Violent fluid movements causing high sideway forces may break rigid probes. If the probe is mounted close to a wall, nozzle or other tank obstruction, noise might appear in the level signal. Therefore the following minimum clearance, according to Table 3-1, must be maintained. Figure 3-2. Free Space Requirement Clearance to tank wall Table 3-1.
Reference Manual Section 3: Installation 00809-0100-4308, Rev BA 3.3.2 December 2014 Flange connection on nozzles Figure 3-3. Mounting in Nozzles Maximum nozzle height Minimum nozzle diameter The transmitter can be mounted in nozzles by using an appropriate flange. It is recommended that the nozzle size is within the dimensions given in Table 3-2. Note The probe must not be in contact with the nozzle (except for the coaxial probe). Table 3-2.
Section 3: Installation Reference Manual December 2014 00809-0100-4308, Rev BA 3.4 Review mounting preparations 3.4.1 Measure tank height The Tank Height is defined as the measured distance from the Upper Reference Point to the Zero Reference Point. Figure 3-4.
Reference Manual Section 3: Installation 00809-0100-4308, Rev BA 3.4.2 December 2014 Shorten the probe In order to leave some clearance distance between the probe end and the tank bottom, the probe might have to be shortened. The goal is to have the probe hang straight so that it does not touch the wall. 2 in. (5 cm) is a suggested value. The probe can be shortened in field. Use the following form to calculate the probe length: Probe Length = Tank height - 2 in.
Section 3: Installation Reference Manual December 2014 00809-0100-4308, Rev BA Flexible single/twin lead probe Note The PTFE covered probes must not be cut in field. Mark where to cut the probe. 2. Remove enough spacers to make place for the weight (only flexible twin lead probes). 3. Loosen the weight. 0 1.
Reference Manual Section 3: Installation 00809-0100-4308, Rev BA Installation December 2014 4. Slide the weight up. 5. Cut the probe at the mark. 6.
Section 3: Installation Reference Manual December 2014 00809-0100-4308, Rev BA Rigid single lead probe Note The PTFE covered probes must not be cut in field. Note Make sure the lead is fixed while cutting. Mark where to cut the probe. 2. Cut the probe at the mark. 0 1.
Reference Manual Section 3: Installation 00809-0100-4308, Rev BA December 2014 Coaxial probe 1. Mark where to cut the probe. 2. Insert the centering piece. (The centering piece is delivered from factory and should be used to prevent the spacers centering the rod from coming loose). 3. Cut the tube to the desired length. 4. Move the centering piece. 5. Cut the rod inside the tube. Make sure the rod is fixed with the centering piece while cutting. Centering piece Pipes longer than 49 in.
Section 3: Installation Reference Manual December 2014 3.4.3 00809-0100-4308, Rev BA Mount a centering disc for pipe installations To prevent the probe from contacting the chamber or pipe wall, centering discs are available for flexible single, rigid single, and flexible twin lead probes. The disc is attached to the end of the probe. Flexible single/twin lead probe Figure 3-6. Centering Disc at the End of the Weight D A B A. Bolt B. Tab washer C. Centering disc D. Weight C 1.
Reference Manual Section 3: Installation 00809-0100-4308, Rev BA December 2014 Rigid single lead probe (8 mm) Note Centering discs shall not be used with PTFE covered probes. 1. Drill one hole using the drilling fixture (included in your shipment). B A A. Drilling fixture B. Probe 2. Mount the bushing, centering disc, and washer at the probe end. Note Do not mount the washer if the centering disc material is PTFE. C A B A. Washer B. Centering disc C. Bushing 3.
Section 3: Installation Reference Manual December 2014 00809-0100-4308, Rev BA 4. Secure the split pin. Rigid single lead probe (13 mm) 1. Drill two holes using the drilling fixture (included in your shipment). B A A. Drilling fixture B. Probe 2. Mount the bushings and centering disc at the probe end. A A B A. Bushing B.
Reference Manual Section 3: Installation 00809-0100-4308, Rev BA 3. December 2014 Adjust distance by shifting hole for split pin in lower bushing. 0.16 in. (4 mm) 0.08 in. (2 mm) Installation 4. Insert the split pins through the bushings and the probe. 5. Secure the split pins.
Section 3: Installation Reference Manual December 2014 3.5 00809-0100-4308, Rev BA Anchor the probe In turbulent tanks it may be necessary to fix the probe. Depending on the probe type, different methods can be used to guide the probe to the tank bottom. This may be needed in order to prevent the probe from hitting the tank wall or other objects in the tank, as well as preventing a probe from breaking. 3.5.
Reference Manual Section 3: Installation 00809-0100-4308, Rev BA December 2014 Figure 3-9. Flexible Twin/Single Lead Probe with Weight and Magnet Magnet 3.5.2 Rigid single lead probe The rigid single lead probe can be guided by a tube welded on the tank bottom. Tubes are customer supplied. Make sure that the probe can move freely in order to handle thermal expansion. The measurement accuracy will be reduced close to the tube opening. Figure 3-10. Rigid Single Lead Probe with Tube Drain 3.5.
Section 3: Installation Reference Manual December 2014 00809-0100-4308, Rev BA Figure 3-11. Coaxial Probe Secured to the Tank Wall 1.1 in. (28 mm) The coaxial probe can be guided by a tube welded on the tank bottom. Tubes are customer supplied. Make sure that the probe can move freely in order to handle thermal expansion. The measurement accuracy will be reduced close to the tube opening. Figure 3-12.
Reference Manual Section 3: Installation 00809-0100-4308, Rev BA 3.6 December 2014 Mount device on tank Mount the transmitter with flange on a nozzle on top of the tank. The transmitter can also be mounted on a threaded connection. Make sure only qualified personnel perform the installation. Note If the transmitter head must be removed from the probe, make sure that the Process seal is carefully protected from dust and water. See “Transmitter head replacement” on page 99 for further information.
Section 3: Installation Reference Manual December 2014 3.6.2 00809-0100-4308, Rev BA 3. Lower the transmitter and probe into the tank. 4. Loosen the nut that connects the transmitter head to the probe slightly. 5. Screw the adapter into the process connection. 6. Rotate the transmitter head so the device display faces the desired direction. 7. Tighten the nut. Max torque is 30 Lbft (40 Nm). 8. Continue with the grounding step. Tank connection with flange Figure 3-14.
Reference Manual Section 3: Installation 00809-0100-4308, Rev BA 3.6.3 December 2014 Tank connection with loose flange (“plate design”) Figure 3-15. Tank Connection with Loose Flange A. Transmitter head B. Nut C. Flange nut D. Bolts E. Probe F. Loose flange G. Plate H. Gasket I. Tank flange A B C D F G E H I The transmitter is delivered with head, flange and probe assembled into one unit.
Section 3: Installation Reference Manual December 2014 3.6.4 00809-0100-4308, Rev BA Tank connection with Tri-Clamp Figure 3-16. Tank connection with Tri-Clamp A B C A. Transmitter head B. Nut C. Tri-Clamp D. Probe E. Clamp F. Gasket G. Tank D E F G 36 1. Place a gasket on top of the tank flange. 2. Lower the transmitter and probe into the tank. 3. Fasten the Tri-Clamp to the tank with a clamp. 4. Loosen the nut that connects the transmitter head to the probe slightly. 5.
Reference Manual Section 3: Installation 00809-0100-4308, Rev BA 3.7 December 2014 Ground the device The Rosemount 3308 Series Wireless Guided Wave Radar transmitter operates with the housing grounded or floating. Floating systems can cause extra noise that may affect many types of readout devices. If the signal appears noisy or erratic, grounding at a single point may solve the problem. Grounding of the electronics enclosure should be done in accordance with local and national installation codes.
Section 3: Installation Reference Manual December 2014 3.8 00809-0100-4308, Rev BA Install the power module Note Use caution when handling the power module, it may be damaged if dropped from heights in excess of 20 ft (6 m). 1. Install the Black Power Module, SmartPower™ Solutions model number 701PBKKF into the transmitter. 2. Close the housing cover and tighten to site or safety specifications.
Reference Manual Section 3: Installation 00809-0100-4308, Rev BA 3.10 December 2014 Utilize the device display If a device display is ordered, it will be shipped attached to the transmitter. The display is ordered in the transmitter model number, option code M5. 3.10.1 Rotate the device display To rotate the display in 90-degree increments, do the following: 1. Squeeze the two black tabs on opposite sides of the display. Refer to Figure 3-20. 2. Gently pull out the display. 3.
Section 3: Installation December 2014 40 Reference Manual 00809-0100-4308, Rev BA Installation
Reference Manual Section 4: Configuration 00809-0100-4308, Rev BA December 2014 Section 4 Configuration Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Safety messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Configuration procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Section 4: Configuration December 2014 4.2 Reference Manual 00809-0100-4308, Rev BA Safety messages Instructions and procedures in this section may require special precautions to ensure the safety of the personnel performing the operations. Information that potentially raises safety issues is indicated by a warning symbol ( ). Please refer to the following safety messages before performing an operation preceded by this symbol. Explosions could result in death or serious injury.
Reference Manual Section 4: Configuration 00809-0100-4308, Rev BA December 2014 Use caution when handling the Power Module. The Power Module may be damaged if dropped from heights in excess of 20 ft (6 m). This device complies with Part 15 of the FCC Rules. Operation is subject to the following conditions: This device may not cause harmful interference. This device must accept any interference received, including interference that may cause undesired operation.
Section 4: Configuration Reference Manual December 2014 4.
Reference Manual Section 4: Configuration 00809-0100-4308, Rev BA December 2014 4.4 Get started with your preferred configuration tool 4.4.1 AMS Wireless Configurator (version 12.0 or later is required) The AMS Wireless Configurator is the recommended software tool for the wireless network devices, and is supplied with the Smart Wireless Gateway. Refer to the AMS Wireless Configurator Manual Supplement (document number 00809-0400-4420) for further information.
Section 4: Configuration Reference Manual December 2014 00809-0100-4308, Rev BA Get the latest Device Descriptor (DD) The Device Descriptor (DD) is a configuration tool that is developed to assist the user through the configuration. The Rosemount 3308 Series DD is typically installed together with AMS Wireless Configurator. To download the latest HART DD, visit the Emerson Process Management Device Install Kit site at: www.emersonprocess.
Reference Manual Section 4: Configuration 00809-0100-4308, Rev BA 4.4.2 December 2014 Field Communicator This section describes how to prepare the Field Communicator to communicate with a Rosemount 3308 Series Transmitter. The Field Communicator can be used to configure the device with a point-to-point connection. Connect the leads on the Field Communicator to the communication terminals of the device as shown in Figure 4-3. Figure 4-3.
Section 4: Configuration Reference Manual December 2014 00809-0100-4308, Rev BA 4.5 Join device to wireless network 4.5.1 Power up the wireless device Make sure that the Smart Wireless Gateway is installed and functioning properly before any wireless field devices are powered. See “Install the power module” on page 38 for further information on how to install the power module. Wireless devices should be powered up in order of proximity from the Gateway, beginning with the closest.
Reference Manual Section 4: Configuration 00809-0100-4308, Rev BA December 2014 7. Secondary Variable 8. Electronics Temperature 9. Supply Voltage Measurement value of mapped Second Variable. Temperature value of device electronics. Voltage reading of the Power Module. 10. Percent of Range 11. Active Alert Present Level value in percent of total measurement range. This screen will only appear if there is at least one Active Alert present.
Section 4: Configuration Reference Manual December 2014 4.5.2 00809-0100-4308, Rev BA Connect to device Connect a Field Communicator or a HART modem to the communication terminals as shown in Figure 4-6. Figure 4-6. Connect to Device Field Communicator 1 4 7 2 5 8 3 6 9 0 HART Modem Communication terminals AMS Wireless Configurator: Field Communicator: a. Start AMS Wireless Configurator. a. Turn on the Field Communicator. b. Select Device Connection View in the View menu. b.
Reference Manual Section 4: Configuration 00809-0100-4308, Rev BA 4.5.3 December 2014 Configure Update Rate The Update Rate is the frequency at which a new measurement is transmitted over the wireless network. The default update rate is 1 minute. This may be changed at commissioning, or at any time via AMS Wireless Configurator or a Field Communicator. The Update Rate is user selectable from 4 seconds to 60 minutes. 1. From the Home screen, select Configure. 2. Go to Guided Setup > Wireless Setup.
Section 4: Configuration Reference Manual December 2014 4.5.4 00809-0100-4308, Rev BA Obtain Network ID and Join Key In order to communicate with the Smart Wireless Gateway, and ultimately the host system, the transmitter must be configured to communicate on the wireless network. This step is the wireless equivalent of connecting wires from a transmitter to the host system.
Reference Manual Section 4: Configuration 00809-0100-4308, Rev BA 4.5.6 December 2014 Verify device joins network Network connection can be verified in four ways, further described in this section: 1. At the device display 2. Using the AMS Wireless Configurator 3. In the Smart Wireless Gateway's integrated web interface 4.
Section 4: Configuration Reference Manual December 2014 00809-0100-4308, Rev BA Figure 4-8. Network Connection Status Screens 54 Network Unknown Network Restarted The device is still in the process of being activated. The device has just restarted. Network Idle The device is starting to join the process. Disconnected from Network Searching for Network Joining the Network The device is in a disconnected state and requires a “Force Join” command to join the network.
Reference Manual Section 4: Configuration 00809-0100-4308, Rev BA December 2014 Verify with AMS Wireless Configurator Start the AMS Wireless Configurator. When the device has joined the network, it will appear in the AMS Wireless Configurator window as illustrated in Figure 4-9. Figure 4-9. AMS Wireless Configurator Screen Figure 4-10 presents the different network connection status images that are shown in the AMS Wireless Configurator Overview screen. Figure 4-10.
Section 4: Configuration Reference Manual December 2014 00809-0100-4308, Rev BA Verify by Gateway To use the Smart Wireless Gateway's integrated web interface, navigate to the Explorer > Status page as shown in Figure 4-11. This page shows whether the device has joined the network and if it is communicating properly. Locate the device in question and verify all status indicators are good (green).
Reference Manual Section 4: Configuration 00809-0100-4308, Rev BA December 2014 4.6 Configure device using Guided Setup 4.6.1 Connect to device Connect to the device using your preferred configuration tool, as shown in Figure 4-12 and Figure 4-13. AMS Wireless Configurator: Field Communicator: a. Start AMS Wireless Configurator. a. Turn on the Field Communicator. b. Select Device Connection View in the View menu. b. Tap the HART symbol from the main menu. c.
Section 4: Configuration Reference Manual December 2014 4.6.2 00809-0100-4308, Rev BA Basic Setup All Basic Setup parameters are described in the section “Configuration parameters” on page 160. Figure 4-14. Guided Setup Screen 4.6.3 1. From the Home screen, select Configure. 2. Go to Guided Setup > Initial Setup. 3. Click Basic Setup, and follow the instructions.
Reference Manual Section 4: Configuration 00809-0100-4308, Rev BA 4.7 December 2014 Verify Level Run the Verify Level tool to match the product level reported by the device to a reference measurement (measured by using for example handgauging). If any difference, the Calibration Offset parameter will be adjusted as shown in Figure 4-15. A minor adjustment using Calibration Offset is normal. There may, for example be a deviation between the actual tank height and the configured value. Non-metallic (e.g.
Section 4: Configuration December 2014 60 Reference Manual 00809-0100-4308, Rev BA Configuration
Reference Manual Section 5: Operation 00809-0100-4308, Rev BA December 2014 Section 5 Operation Safety messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Device display screen messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . View measurement values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Check Device Status . . .
Section 5: Operation December 2014 Reference Manual 00809-0100-4308, Rev BA Process leaks could result in death or serious injury. Handle the transmitter carefully. If the process seal is damaged, gas could escape from the tank when removing the transmitter head from the probe. Only qualified personnel should install the equipment. Do not remove the transmitter while in operation.
Reference Manual Section 5: Operation 00809-0100-4308, Rev BA 5.2 December 2014 Device display screen messages The device display can be used to present different variables and a diagnostic screen sequence. 5.2.1 Variable screens If the Display Mode is set to Periodic, the device display shows a periodic sequence of user-chosen variables during operation. A new screen appears according to configured wireless update rate. The device display will also show ALERT PRESNT if at least one alert is present.
Section 5: Operation Reference Manual December 2014 5.2.2 00809-0100-4308, Rev BA Diagnostic button screen sequence The diagnostic button screen sequence on the device display can be used to obtain detailed diagnostic information. Do the following: 64 1. Unscrew the device display cover. 2. Press and hold the “DIAG” button until the first diagnostic button screen appears on the device display. Then release the “DIAG” button. 3.
Reference Manual Section 5: Operation 00809-0100-4308, Rev BA December 2014 Figure 5-1. Diagnostic Button Screen Sequence Operation 1. Device Information: Tag 2. Device Serial Number 3. Software Revision User entered tag which is 8 characters long. This screen will not display if all characters are blank. Used to determine Device Serial Used to determine Device Number. Software Revision. 4. Network ID 5. Network Connection Status 6.
Section 5: Operation Reference Manual December 2014 5.3 00809-0100-4308, Rev BA View measurement values Measurement values can be viewed using AMS Wireless Configurator and Field Communicator. 5.3.1 View current measurement values Current measurement data of the Primary Variable (PV) and Secondary Variable (SV) are presented on the Overview screen. To view all current measurement values, do the following: 5.3.2 1. From the Home Screen, go to Service Tools > Variables. 2.
Reference Manual Section 5: Operation 00809-0100-4308, Rev BA 5.4 December 2014 Check Device Status The overall device status is presented in AMS Wireless Configurator and Field Communicator under the Overview screen. The Rosemount 3308 Series reports diagnostic alerts when there is a device malfunction. For information on these alerts, see “Alert messages in AMS Wireless Configurator and Field Communicator” on page 74.
Section 5: Operation December 2014 68 Reference Manual 00809-0100-4308, Rev BA Operation
Reference Manual Section 6: Service and Troubleshooting 00809-0100-4308, Rev BA December 2014 Section 6 Service and Troubleshooting Safety messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Alert messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Troubleshooting guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Section 6: Service and Troubleshooting December 2014 Reference Manual 00809-0100-4308, Rev BA Electrical shock can result in death or serious injury. Ground device on non-metallic tanks (e.g. fiberglass tanks) to prevent electrostatic charge build-up. Single lead probes are sensitive for strong electromagnetic fields and therefore not suitable for non-metallic tanks. Care must be taken during transportation of power module to prevent electrostatic charge build-up.
Reference Manual Section 6: Service and Troubleshooting 00809-0100-4308, Rev BA December 2014 Use caution when handling the Power Module. The Power Module may be damaged if dropped from heights in excess of 20 ft (6 m). This device complies with Part 15 of the FCC Rules. Operation is subject to the following conditions: This device may not cause harmful interference. This device must accept any interference received, including interference that may cause undesired operation.
Section 6: Service and Troubleshooting December 2014 Reference Manual 00809-0100-4308, Rev BA 6.2 Alert messages 6.2.1 Device display alerts The following active alert screens will show the device diagnostics depending on the state of the device. If the device display shows ALERT PRESNT but none of the following screens appear, then go to the Active Alerts screen in AMS® Wireless Configurator or Field Communicator for further information.
Reference Manual Section 6: Service and Troubleshooting 00809-0100-4308, Rev BA December 2014 Figure 6-1. Active Alerts Screens Bandwidth Limited Configuration Warning Configuration Failure Electronics Warning The device has not yet received all of the requested wireless bandwidth needed to operate as configured. See “Wireless Network troubleshooting” on page 82 for recommended actions. The device has detected a configuration error. Non-critical operation of the device may be affected.
Section 6: Service and Troubleshooting Reference Manual December 2014 6.2.2 00809-0100-4308, Rev BA Alert messages in AMS Wireless Configurator and Field Communicator Table 6-1 to Table 6-4 shows list of alert messages that may be displayed in the AMS Wireless Configurator and Field Communicator. To view Active Alerts, do the following: From the Home Screen, go to Service Tools > Active Alerts. Table 6-1.
Reference Manual Section 6: Service and Troubleshooting 00809-0100-4308, Rev BA December 2014 Table 6-2. Configuration Error Details (D:) Message Description Recommended actions Lower Range Value or Upper Range Value is out of limits The lower/upper range value is outside the lower/upper sensor limits. This is outside the range where the sensor works properly, hence the measurement may be unreliable. 1. Check the lower and upper range values in relation to the sensor limits.
Section 6: Service and Troubleshooting December 2014 Reference Manual 00809-0100-4308, Rev BA Table 6-3. Maintenance Alerts (M:) Message Description Supply Voltage Low The supply voltage is low and may affect Device Operation. 1. Replace the Power Module, see “Power module replacement” on page 97. Electronics Temperature Out of Limits The temperature of the electronics board has exceeded the transmitter’s operating range. 1. Verify that ambient temperature is within the operating range. 2.
Reference Manual Section 6: Service and Troubleshooting 00809-0100-4308, Rev BA December 2014 Table 6-4. Advisory Alerts (A:) Message Description Database Memory Warning The device has failed to write to the database memory at some time in the past. Any data written during this time may have been lost. 1. If logging dynamic data is not needed, this advisory alert can be safely ignored. 2. Restart the device. 3. Reconfirm all configuration items in the device. 4.
Section 6: Service and Troubleshooting Reference Manual December 2014 6.3 00809-0100-4308, Rev BA Troubleshooting guide If there is a malfunction despite the absence of alerts, see Table 6.3.1 for information on possible causes and recommended actions. The troubleshooting guide contains the following symptoms: 6.3.1 Incorrect level readings, see page 78. Incorrect or missing interface level reading, see page 80. Power module troubleshooting, see page 81.
Reference Manual Section 6: Service and Troubleshooting 00809-0100-4308, Rev BA Symptom December 2014 Possible cause and recommended actions The tank is full. Check the product level. Check Thresholds, see “Adjusting thresholds” on page 84. The transmitter has locked on disturbing obstacles at top of the tank. See “Handling disturbances at the top of the tank” on page 92 for recommended actions. The transmitter is configured with wrong Probe Type, refer to “Probe Type” on page 161.
Section 6: Service and Troubleshooting Reference Manual December 2014 6.3.2 00809-0100-4308, Rev BA Incorrect or missing interface level reading Symptom The Upper Product Dielectric Constant is not correct, see “Upper Product Dielectric Constant” on page 163. Air gap is too big when the Measurement Mode is set to Interface Level with Submerged Probe. Reduce air gap or switch the Measurement Mode to Product Level and Interface Level. The Measurement Mode is set to Product Level.
Reference Manual Section 6: Service and Troubleshooting 00809-0100-4308, Rev BA 6.3.3 Power module troubleshooting Symptom The Power Module seems to run out of battery very fast. 6.3.4 December 2014 Possible cause and recommended actions Consider Update Rate, refer to “Configure Update Rate” on page 51. Selecting a fast update rate has an impact on Power Module life. Consider Performance Mode, refer to “Performance Mode” on page 167.
Section 6: Service and Troubleshooting Reference Manual December 2014 6.3.5 00809-0100-4308, Rev BA Wireless Network troubleshooting Symptom The device is not joining the wireless network. Limited Bandwidth error 82 Possible cause and recommended actions Verify Network ID and Join Key. The Network ID and Join Key in the device must match the Network ID and Join Key of the Gateway, refer to “Obtain Network ID and Join Key” on page 52 and “Enter Network ID and Join Key” on page 52.
Reference Manual Section 6: Service and Troubleshooting 00809-0100-4308, Rev BA 6.4 December 2014 Service and troubleshooting tools This section briefly describes tools and functions in the AMS Wireless Configurator and Field Communicator which may be useful for service and troubleshooting of Rosemount 3308 Series Transmitter. 6.4.1 Reading the Echo Curve The AMS Wireless Configurator and Field Communicator have powerful tools for advanced troubleshooting.
Section 6: Service and Troubleshooting Reference Manual December 2014 6.4.2 00809-0100-4308, Rev BA Adjusting thresholds Measurement with the Rosemount 3308 is based on the fact that the radar signal pulses are reflected by the product surface and the interface between two liquids. Signal amplitude thresholds are used to separate the measurement signal from disturbing echoes and noise.
Reference Manual Section 6: Service and Troubleshooting 00809-0100-4308, Rev BA December 2014 Guidelines for setting the Surface Threshold Before changing the Surface Threshold, make sure the product level is at least 20 in. (0.5 m) from the lower side of the device flange. Set the Surface Threshold to about 1/3 of the weakest surface echo amplitude in the measuring range.
Section 6: Service and Troubleshooting December 2014 Reference Manual 00809-0100-4308, Rev BA Guidelines for setting the Interface Threshold The Interface Threshold should be approximately 50% of the interface signal amplitude. If possible, Interface Threshold should be higher than Surface Threshold. Procedure for adjusting thresholds To adjust the amplitude thresholds, do the following: 1. 2. Read the Echo Curve: a. Start the Echo Curve reading, refer to “Reading the Echo Curve” on page 83.
Reference Manual Section 6: Service and Troubleshooting 00809-0100-4308, Rev BA December 2014 Figure 6-4. Surface Threshold Is Too High 100 250 80 Surface Threshold is above the Product Surface peak AMPLITUDE 60 40 200 150 20 0 100 -20 50 -40 -60 0 DISTANCE If there are disturbing objects in the tank, the Surface Threshold must be carefully set to avoid locking on the wrong amplitude peak.
Section 6: Service and Troubleshooting Reference Manual December 2014 00809-0100-4308, Rev BA By adjusting the Surface Threshold the product surface is properly detected as illustrated in Figure 6-6: Figure 6-6.
Reference Manual Section 6: Service and Troubleshooting 00809-0100-4308, Rev BA December 2014 By adjusting Interface Threshold, the peak at the interface between the upper and lower products is detected as illustrated in Figure 6-8. AMPLITUDE Figure 6-8. After Changing the Interface Threshold the Transmitter Detects the Interface After Interface Threshold is adjusted the interface is correctly detected Surface Threshold DISTANCE 6.4.
Section 6: Service and Troubleshooting Reference Manual December 2014 00809-0100-4308, Rev BA Network Join Details Obtain detailed information about the network join, and configure how the device attempts to join the network. From the Home Screen, go to Service Tools > Communications > Join Details. Join Details includes: Join Mode 6.4.5 This mode configures how the device attempts to join the network.
Reference Manual Section 6: Service and Troubleshooting 00809-0100-4308, Rev BA 6.4.6 December 2014 Using the Simulation Mode This function can be used to simulate measurements and alerts. 1. From the Home Screen, go to Service Tools > Simulate. 2. Click desired variable and follow the on-screen instructions. 6.5 Application challenges 6.5.1 Resolving thin oil layers Thin oil layer on top of water might cause jumpy surface readings.
Section 6: Service and Troubleshooting Reference Manual December 2014 00809-0100-4308, Rev BA To set the Peak Detection Method to Threshold Intersection, do the following: 1. From the Home Screen, go to Service Tools > Echo Tuning > Advanced. 2. Under Peak Detection Method, select Threshold Intersection and then click Send. 3. Run Verify Level to compensate for any offset error (see Figure 6-12) introduced by the Threshold Intersection method. See “Verify Level” on page 59 for further instructions.
Reference Manual Section 6: Service and Troubleshooting 00809-0100-4308, Rev BA December 2014 Figure 6-13. Near Zone Threshold Reference Peak Disturbance Near Zone Threshold blocking the disturbance Surface Threshold Product Surface Peak Guidelines for setting the Near Zone Threshold The Near Zone Threshold must be higher than the Surface Threshold to have an effect. The threshold in the near zone is set to the highest value of the configured Near Zone Threshold and Surface Threshold.
Section 6: Service and Troubleshooting Reference Manual December 2014 00809-0100-4308, Rev BA Using the Trim Near Zone function The Trim Near Zone function is used to fine tune performance in the area close to the tank top. Normally it is not necessary to use the function, but if you experience problems related to the nozzle, pipe, or chamber installation, you may need to use this function. Figure 6-14 describes the Trim Near Zone function and its effect on the echo curve.
Reference Manual Section 6: Service and Troubleshooting 00809-0100-4308, Rev BA December 2014 Note Before changing the Upper Null Zone, check entered limit and deadband values for the High Level Alerts. High Level Alerts should not be placed in the Upper Null Zone. To set the Upper Null Zone do the following: 1. Identify desired Upper Null Zone using the Echo Curve plot: 2. a. Start the Echo Curve reading, refer to “Reading the Echo Curve” on page 83. b.
Section 6: Service and Troubleshooting Reference Manual December 2014 6.5.3 00809-0100-4308, Rev BA Interface measurements with fully submerged probes The Measurement Mode Interface Level with Submerged Probe is used to handle interface measurements when the product level is not visible, for example in a full chamber pipe as illustrated in Figure 6-17. In this case the probe is fully submerged into the upper product, and only the interface level is detected by the transmitter.
Reference Manual Section 6: Service and Troubleshooting 00809-0100-4308, Rev BA 6.5.4 December 2014 Noise or weak surface echoes In order to increase the measurement performance in difficult applications where the surface echo peak is low compared to the noise, it is recommended to set the Performance Mode to High (Short battery life). A low surface peak compared to the noise might be caused by a turbulent surface, low dielectric constant, plastic tanks and so on.
Section 6: Service and Troubleshooting December 2014 Reference Manual 00809-0100-4308, Rev BA Handling considerations The Black Power Module with the wireless unit contains two "C" size primary lithium/thionyl chloride batteries. Each battery contains approximately 2.5 grams of lithium, for a total of 5 grams in each Power Module. Under normal conditions, the battery materials are self-contained and are not reactive as long as the batteries and the battery pack integrity are maintained.
Reference Manual Section 6: Service and Troubleshooting 00809-0100-4308, Rev BA 6.7 December 2014 Transmitter head replacement Figure 6-18. Transmitter Head Replacement Flange version Threaded version A A Put the protection plug here! B Put the protection plug here! B C A. Nut B. Process Seal C. Adapter 1. Loosen the nut that connects the transmitter head to the process seal. 2. Carefully lift the transmitter head. 3.
Section 6: Service and Troubleshooting Reference Manual December 2014 6.8 00809-0100-4308, Rev BA Probe replacement Figure 6-19. Probe Replacement Flange version Threaded version A A B B C C D D A. Transmitter head B. Nut C. Process Seal D. Probe 1. Loosen the nut. 2. Remove the transmitter head from the old probe. Make sure to protect the transmitter head bottom from dust and water. 3.
Reference Manual Appendix A: Reference Data 00809-0100-4308, Rev BA December 2014 Appendix A Reference Data Functional specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 101 Performance specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 107 Physical specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Appendix A: Reference Data December 2014 Reference Manual 00809-0100-4308, Rev BA Radio frequency output from antenna External (WK option) antenna: Maximum10 mW (+10dBm) EIRP High Gain, Remote (WN option) antenna: Maximum of 40mW (16dBm) EIRP Modulation type QPSK/iEEE 802.15.4 DSSS IEC 62591 (WirelessHART) Number of channels 15 Channel spacing 5 MHz Emission designation G1D Transmit rate User selectable, 4 seconds to 60 minutes A.1.
Reference Manual Appendix A: Reference Data 00809-0100-4308, Rev BA December 2014 Output variables Display PV, SV, TV, QV Level X X Distance X X Surface Signal Strength N/A X(2) Total Volume X X X X X X N/A X(2) Upper Product Thickness(3) X X Electronics Temperature X X(2) Signal Quality X X(2) Supply Voltage X X(2) % of Range X X(2) Interface Level (1) Interface Distance(1) Interface Signal Strength (1) (1) For 3308Axx1, Interface measurement is only available for
Appendix A: Reference Data Reference Manual December 2014 A.1.5 00809-0100-4308, Rev BA Process temperature and pressure rating Process temperature Figure A-2. Max. Rating, Standard Tank Connections Pressure psig (bar) 580 (40) PTFE covered probe and flange (model code 7) 232 (16) -14 (-1) -40 (-40) 302 (150) Temperature °F (°C) Note The maximum process temperature is at the lower part of the flange. Final rating depends on flange and O-ring selection.
Reference Manual Appendix A: Reference Data 00809-0100-4308, Rev BA December 2014 Fisher® & Masoneilan® flange rating 316L SST Flanges according to ASME B16.5 Table 2-2.3: Max. 302 °F/580 psig (150 °C/40 bar) JIS flange rating 316L SST Flanges according to JIS B2220 material group 2.3: Max. 302 °F/580 psig (150 °C/40 bar) Tri-Clamp™ rating Maximum pressure is 16 bar for 1½-in. (37.5 mm) and 2-in. (50 mm) housing; and 10 bar for 3-in. (75 mm) and 4-in. (100 mm) housing.
Appendix A: Reference Data Reference Manual December 2014 A.1.6 00809-0100-4308, Rev BA Interface measurements The Rosemount 3308 Series is well suited for interface measurements, including applications where the probe is fully submerged in the liquid. Figure A-3.
Reference Manual Appendix A: Reference Data 00809-0100-4308, Rev BA December 2014 A.2 Performance specifications A.2.1 General Reference conditions(1) Single flexible probe mounted in a 4" pipe. Normal indoor temperature (68° - 79 °F, 20° - 26 °C) water. Reference accuracy Accuracy ±0.98 in. (25 mm) ±0.25 in. (6 mm) 0 0 33 ft. (10 m) 56 ft. (17 m) Measuring distance Repeatability ±0.08 in. (2 mm)(2) Ambient temperature effect Less than 0.
Appendix A: Reference Data Reference Manual December 2014 A.2.3 00809-0100-4308, Rev BA Accuracy over measuring range The measuring range depends on probe type, dielectric constant of the product and installation environment, and is limited by the Blind Zones at the very top and bottom of the probe. In the Blind Zones, the accuracy exceeds ±1.18 in. (30 mm), and measurements may not be possible. Measurements close to the Blind Zones will have reduced accuracy.
Reference Manual Appendix A: Reference Data 00809-0100-4308, Rev BA December 2014 Figure A-5. Accuracy over Measuring Range for Flexible Twin Lead Probe Water (DC = 80) Oil (DC = 2) ±0.25 in. ±1.18 in. ±0.25 in. ±1.18 in. (6 mm) (30 mm) (6 mm) (30 mm) 3.2 in. (8 cm) 3.2 in. (8 cm) 7.9 in. (20 cm) 7.9 in. (20 cm) Blind Zone 0.4 in. (1 cm) 2 in. (5 cm) 7.9 in. (20 cm) 2.8 in. (7 cm) Figure A-6. Accuracy over Measuring Range for Coaxial Probe Water (DC = 80) Oil (DC = 2) ±0.25 in. ±1.
Appendix A: Reference Data Reference Manual December 2014 A.2.4 00809-0100-4308, Rev BA Maximum measuring range and minimum dielectric constant Table A-3. Measuring Range and Minimum Dielectric Constant Flexible single lead Rigid single lead Flexible twin lead Coaxial 55.8 ft (17 m) 19.7 ft (6 m) 9.8 ft (3 m) for 8 mm probes (code 4A) Maximum measuring range 55.8 ft (17 m) 19.7 ft (6 m) for 13 mm probes (code 4B) Minimum dielectric constant(1)(2) 2.0 up to 32.8 ft (10 m) 2.0 10 up to 55.
Reference Manual Appendix A: Reference Data 00809-0100-4308, Rev BA A.2.6 December 2014 Viscosity and Coating/Build-up Table A-4. Maximum Recommended Viscosity and Coating/Build-up Single lead Flexible twin lead Coaxial Maximum viscosity 8000 cP(1) 1500 cP 500 cP Coating / build-up Coating allowed Thin coating allowed, but no bridging Coating not recommended (1) Consult your local Emerson Process Management representative in the case of agitation/turbulence and high viscous products. A.
Appendix A: Reference Data Reference Manual December 2014 00809-0100-4308, Rev BA Flange dimensions Follows ASME B16.5, JIS B2220, and EN 1092-1 standards for blind flanges. For Proprietary Fisher and Masoneilan flanges, see “Proprietary Flanges” on page 137. Probe versions Flexible Single Lead, Rigid Single Lead, Flexible Twin Lead, and Coaxial. There are in total four weight and anchoring options for Flexible Single Lead probes. Table A-5.
Reference Manual Appendix A: Reference Data 00809-0100-4308, Rev BA December 2014 Total probe length This is defined from the Upper Reference Point to the end of the probe (weight included, if applicable). Figure A-9. Total Probe Length NPT BSP/G Flange Tri-Clamp Upper Reference Point Total Probe Length Select the probe length according to the required measuring range (the probe must be hung and fully extended through the entire distance where level readings are desired).
Appendix A: Reference Data Reference Manual December 2014 00809-0100-4308, Rev BA Collapse load Flexible Single Lead: 3597 lb (16 kN) Sideway capacity Rigid Single Lead: 4.4 ft. lbf, 0.44 lb at 9.8 ft. (6 Nm, 0.2 kg at 3 m) Coaxial: 73.7 ft. lbf, 3.7 lb at 19.7 ft. (100 Nm, 1.67 kg at 6 m) Minimum pipe/bypass diameter Flexible Single Lead: Consult your local Emerson Process Management representative. Rigid Single Lead: 2 in.
Reference Manual Appendix A: Reference Data 00809-0100-4308, Rev BA A.3.3 December 2014 Engineered solutions When standard model codes are not sufficient to fulfill requirements, please consult the factory to explore possible Engineered Solutions. This is typically, but not exclusively, related to the choice of wetted materials or the design of a process connection. These Engineered Solutions are part of the expanded offerings and may be subject to additional delivery lead time.
Appendix A: Reference Data Reference Manual December 2014 00809-0100-4308, Rev BA The probe length to use for a Rosemount 9901 chamber can be calculated with this formula: Side-and-side dimension: Probe length=Centre-to-centre dimension + 19 in. (48 cm) Side-and-bottom dimension: Probe length=Centre-to-centre dimension + 4 in. (10 cm) Use a centering disc the same diameter as the chamber if the probe length >3.3 ft. (1 m). See “Centering discs” on page 118 for which disc to use.
Reference Manual Appendix A: Reference Data 00809-0100-4308, Rev BA December 2014 Table A-6. Required Probe Length in Chambers Chamber manufacturer Probe length(1) Major torque-tube manufacture (249B, 249C, 249K, 249N, 259B) Displacer + 9 in. (229 mm) Masoneilan (Torque tube operated), proprietary flange Displacer + 8 in. (203 mm) Other - torque tube(2) Displacer + 8 in. (203 mm) (3) Magnetrol (spring operated) Displacer + between 7.8 in. (195 mm) to 15 in.
Appendix A: Reference Data Reference Manual December 2014 00809-0100-4308, Rev BA Centering discs To prevent the probe from contacting the chamber or pipe wall, centering discs are available for rigid single, flexible single, and flexible twin lead probes. The disc is attached to the end of the probe. Discs are made of stainless steel or PTFE. See Table A-7 on page 118 for Dimension D. Table A-8 on page 118 shows which centering disc diameter to choose for a particular pipe. Figure A-12.
Reference Manual Appendix A: Reference Data 00809-0100-4308, Rev BA A.4 December 2014 Ordering information Specification and selection of product materials, options, or components must be made by the purchaser of the equipment. See page 111 for more information on Material Selection. Table A-9. 3308 Series Level and/or Interface Measurements in Liquids Ordering Information The starred options (★) represent the most common options and should be selected for best delivery.
Appendix A: Reference Data Reference Manual December 2014 00809-0100-4308, Rev BA Table A-9. 3308 Series Level and/or Interface Measurements in Liquids Ordering Information The starred options (★) represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Reference Manual Appendix A: Reference Data 00809-0100-4308, Rev BA December 2014 Table A-9. 3308 Series Level and/or Interface Measurements in Liquids Ordering Information The starred options (★) represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Appendix A: Reference Data December 2014 Reference Manual 00809-0100-4308, Rev BA Table A-9. 3308 Series Level and/or Interface Measurements in Liquids Ordering Information The starred options (★) represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Reference Manual Appendix A: Reference Data 00809-0100-4308, Rev BA December 2014 Table A-9. 3308 Series Level and/or Interface Measurements in Liquids Ordering Information The starred options (★) represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time. Engineered solutions (see page 115) Pxxx (1) (2) (3) (4) (5) Engineered Solutions beyond standard model codes.
Appendix A: Reference Data December 2014 A.5 Reference Manual 00809-0100-4308, Rev BA Spare parts and accessories Table A-11.
Reference Manual Appendix A: Reference Data 00809-0100-4308, Rev BA December 2014 Table A-11.
Appendix A: Reference Data Reference Manual December 2014 00809-0100-4308, Rev BA Table A-12.
Reference Manual Appendix A: Reference Data 00809-0100-4308, Rev BA December 2014 Table A-12. 3308 Series Spare Parts List - Probe Process connection rating (see Table A-10 on page 123 for availability) NN For use with non-flange process connection type ASME rating AA ASME B16.5 Class 150 Flange AB ASME B16.
Appendix A: Reference Data Reference Manual December 2014 00809-0100-4308, Rev BA Table A-12. 3308 Series Spare Parts List - Probe 2A Flexible Twin Lead with weight Flange / 1½-in., 2-in. Thread Min.: 3 ft. 4 in. (1 m) Max.: 55 ft. 9 in. (17 m) 3A Coaxial (for level measurement)(2) Flange / 1-in., 1½-in., 2-in. Thread Min.: 1 ft. 4 in. (0.4 m) Max.: 19 ft. 8 in.
Reference Manual Appendix A: Reference Data 00809-0100-4308, Rev BA December 2014 Table A-12. 3308 Series Spare Parts List - Probe Assemble / consolidate to chamber (see page 115) XC Consolidate to Chamber Engineered solutions (see page 115) Pxxx (1) (2) (3) (4) Engineered Solutions beyond standard model codes. (Consult factory for details) Minimum probe length is 4 ft 11 in. (1.5 m) for PTFE covered probes (Material of Construction codes 7 and 8). Requires model 3308Axx1.
Appendix A: Reference Data Reference Manual December 2014 00809-0100-4308, Rev BA Table A-13. Accessories Ordering Information The starred options (★) represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time. Process connection - size/type (consult the factory for other process connections) Centering discs(1)(2) (see page 118 for dimensions and size recommendation) Outer diameter 03300-1655-1001 Kit, 2-in.
Reference Manual Appendix A: Reference Data 00809-0100-4308, Rev BA A.6 December 2014 Dimensional drawings Figure A-13. Flexible Single Lead Probe 2.8 (71.3) Dimensions are in inches (millimeters). 3.9 (100.2) External antenna (option WK) NPT 1 in., s52 NPT 1½ in., s52 NPT 2 in., s60 12 (297.2) Thread sealing 2.4 (62) H (Nozzle height) H < 4 (100) + D Ø D (Nozzle diameter) 1.9 (47) 3.4 (86) Recommended: D > 4 (100) Minimum: D=1.5 (38)(1) Ø 0.16 (4): SST probe Ø 0.
Appendix A: Reference Data Reference Manual December 2014 00809-0100-4308, Rev BA Figure A-14. Rigid Single Lead Probe Dimensions are in inches (millimeters). 2.8 (71.3) 3.9 (100.2) External antenna (option WK) 12 (297.2) NPT 1 in., s52 NPT 1½ in., s52 NPT 2 in., s60 Thread sealing 2.4 (62) Nozzle height (H) H < 4 (100) + D Ø D (Nozzle diameter) 1.9 (47) 3.4 (86) Recommended: D > 4 (100) Minimum: D=1.5 (38) for probe type 4A(1) D=2 (50) for probe type 4B(1) L 10 ft (3 m) for Ø 0.31 (8) Min.
Reference Manual Appendix A: Reference Data 00809-0100-4308, Rev BA December 2014 Figure A-15. Coaxial Probe Dimensions are in inches (millimeters). 2.8 (71.3) 3.9 (100.2) External antenna (option WK) NPT 1 in., s52 NPT 1½ in., s52 NPT 2 in., s60 12 (297.2) Thread sealing 1.9 (47) Nozzle diameter Min. Ø 1.2 (30) 2.4 (62) 3.4 (86) Min. 8 (200) L 20 ft. (6 m) Metallic contact with nozzle or tank allowed Min. 0 to tank wall Chamber/pipe Ø 1.1 (28) Min. 0.2 (5) Min. Ø 1.3 (32) 7.
Appendix A: Reference Data Reference Manual December 2014 00809-0100-4308, Rev BA Figure A-16. Flexible Twin Lead Probe Dimensions are in inches (millimeters). 2.8 (71.3) 3.9 (100.2) External antenna (option WK) 12 (299.5) NPT 1½ in., s52 NPT 2 in., s60 Thread sealing 1.1 (27) Nozzle height (H) H < 4 (100) + D 1.8 (45) 3.5 (88) Ø D (Nozzle diameter) Recommended: D > 4 (100) Minimum: D=2 (50)(1) Chamber/pipe L 56 ft. (17 m) Min. 8 (200) Ø 0.16 (4) Ø 0.16 (4) Min.
Reference Manual Appendix A: Reference Data 00809-0100-4308, Rev BA December 2014 Figure A-17. Mounting Bracket (Option Code BR) IVE ATMOS OS P PL RE HE IN E X Dimensions are in inches (millimeters). AL EP IV E KE T IG HT W H E N CI R C UI T Pipe diameter max 2.5 (64) 5.2 (133) Pipe mounting (vertical pipe) Pipe mounting (horizontal pipe) 2.2 (57) 0.3 (7) 2.8 (70) 0.
Appendix A: Reference Data Reference Manual December 2014 00809-0100-4308, Rev BA Figure A-18. High Gain, Remote Antenna (Option Code WN1) Dimensions are in inches (millimeters). Mounting bracket 0.53 (13.5) 2.75 (69.9) 0.50 (12.7) Antenna 20.2 (513) 2.75 (69.9) 4.22 (107.2) 0.13 (3.3) Ø 0.41 (10.4) 3.80 (96.5) Mounting bracket 1.05 (26.7) Ø 0.66 (16.7) 2.50 (63.5) 25 ft (7.
Reference Manual Appendix A: Reference Data 00809-0100-4308, Rev BA December 2014 Figure A-19. Proprietary Flanges Raised Face Recessed Face Dimensions are in inches (millimeters).
Appendix A: Reference Data December 2014 138 Reference Manual 00809-0100-4308, Rev BA Reference Data
Reference Manual Appendix B: Product Certifications 00809-0100-4308, Rev BA December 2014 Appendix B Product Certifications Safety messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 139 Product certifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 141 Approval drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Appendix B: Product Certifications December 2014 Reference Manual 00809-0100-4308, Rev BA Electrical shock can result in death or serious injury. Ground device on non-metallic tanks (e.g. fiberglass tanks) to prevent electrostatic charge build-up. Single lead probes are sensitive for strong electromagnetic fields and therefore not suitable for non-metallic tanks. Care must be taken during transportation of power module to prevent electrostatic charge build-up.
Reference Manual Appendix B: Product Certifications 00809-0100-4308, Rev BA B.2 Product certifications B.2.1 European Union Directive Information December 2014 The EC Declaration of Conformity for all applicable European directives for this product can be found in the 3308A Product Certifications document (document number 00825-0200-4308). The most current revision is available at www.rosemount.com. A hard copy may be obtained by contacting your local sales representative. B.2.
Appendix B: Product Certifications December 2014 B.2.7 Reference Manual 00809-0100-4308, Rev BA FCC and IC This device complies with Part 15 of the FCC Rules. Operation is subject to the following conditions: This device may not cause harmful interference and this device must accept any interference, including any interference that may cause undesired operation of the device. This device must be installed to ensure a minimum antenna separation distance of 20 cm from all persons.
Reference Manual Appendix B: Product Certifications 00809-0100-4308, Rev BA B.2.10 December 2014 Hazardous Locations Certificates North American Certifications US Approvals I5 Intrinsically Safe Certificate No: 3046655 Applicable Standards: FM Class 3600 – 2011, FM Class 3610 – 2010, FM Class 3810 – 2005, NEMA 250 – 2003, ANSI/ISA 60079-0:2009, ANSI/ISA 60079-11:2011, ANSI/ISA 60079-26:2011, ANSI/ISA 60529:2004.
Appendix B: Product Certifications December 2014 Reference Manual 00809-0100-4308, Rev BA Canadian Approval I6 Intrinsically Safe Certificate No: 3046655 Applicable Standards: CSA Std. C22.2 No. 1010.1:04, CSA Std. 22.2 No 94-M91, CSA Std. C22.2 No. 157 – 92, CAN/CSA-C22.2 No. 60079-0:11, CAN/CSA-C22.2 No. 60079-11:11 Markings: INTRINSICALLY SAFE Ex ia CLASS I, GP A, B, C, D; CLASS I, Zone 0, Ex ia IIC Ga; TEMP CODE T4 (-55 °C Ta +70 °C) WHEN INSTALLED PER ROSEMOUNT DRAWING 03308-1010.
Reference Manual Appendix B: Product Certifications 00809-0100-4308, Rev BA December 2014 European Certifications I1 ATEX Intrinsic Safety Certificate No: FM 12ATEX0072X Applicable Standards: EN 60079-0:2012, EN 60079-11: 2012, EN 60079-26:2007 Markings: Category II 1 G, Ex ia IIC T4 Ga (-55 °C Ta +70 °C); 1180 Special Conditions of Certification: Product Certifications 1.
Appendix B: Product Certifications December 2014 Reference Manual 00809-0100-4308, Rev BA IECEx Certifications I7 IECEx Intrinsic Safety Certificate No: IECEx FMG 12.0029X Applicable Standards: IEC 60079-0: 2011, IEC 60079-11: 2011, IEC 60079-26:2006 Markings: Ex ia IIC T4 Ga (-55 °C Ta +70 °C) Special Conditions of Certification: 146 1. The Model 3308 transmitter housing contains aluminum, protect the enclosure to avoid a potential risk of ignition due to impact or friction. 2.
Reference Manual Appendix B: Product Certifications 00809-0100-4308, Rev BA December 2014 Taiwan Certifications 注意! 依據 低功率電波輻射性電機管理辦法 第十二條 經型式認證合格之低功率射頻電機,非經許可,公司、商號或使用者均不 得擅自變更頻率、加大功率或變更原設計之特性及功能。 第十四條 低功率射頻電機之使用不得影響飛航安全及干擾合法通信;經發現有干擾 現象時,應立即停用,並改善至無干擾時方得繼續使用。 前項合法通信,指依電信法規定作業之無線電通信。 低功率射頻電機須忍受合法通信或工業、科學及醫療用電波輻射性電機設備 之干擾。 B.2.11 Other certifications U1 Overfill protection Certificate: Z-65.
Appendix B: Product Certifications December 2014 B.3 Reference Manual 00809-0100-4308, Rev BA Approval drawings This section contains Factory Mutual installation drawings. The installation guidelines must be followed to maintain certified ratings for installed transmitters.
Reference Manual 00809-0100-4308, Rev BA Appendix B: Product Certifications December 2014 Electronic Master – PRINTED COPIES ARE UNCONTROLLED – Rosemount Proprietary Figure B-1.
Appendix B: Product Certifications December 2014 150 Reference Manual 00809-0100-4308, Rev BA Product Certifications
Reference Manual Appendix C: High Gain Remote Antenna Option 00809-0100-4308, Rev BA December 2014 Appendix C High Gain Remote Antenna Option Safety messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Functional and physical specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Review installation considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Appendix C: High Gain Remote Antenna Option Reference Manual December 2014 00809-0100-4308, Rev BA C.2 Functional and physical specifications C.2.1 General C.2.2 C.2.3 C.2.4 C.2.5 C.2.6 152 Weight: 1.0 lb (0.4 kg Ratings: NEMA 4X, and IP66/67 Vibration: 3g Max vibration Wireless Output: WirelessHART® 2.4 GHz DSSS (Direct Sequence Spread Spectrum) Communication range: 2/3 mile (3,300 feet) (1.0 km) with L.O.S.
Reference Manual Appendix C: High Gain Remote Antenna Option 00809-0100-4308, Rev BA December 2014 C.3 Review installation considerations C.3.1 Antenna mounting Mount antenna vertically (±5°) C.3.2 Antenna height Mount antenna 14 feet (4.3 meters) above infrastructure with clear line of sight. C.3.3 Affix coaxial cable Ensure that coaxial cable is securely affixed to the mast to avoid excessive cable movement. C.3.
Appendix C: High Gain Remote Antenna Option Reference Manual December 2014 C.5 00809-0100-4308, Rev BA Install the high gain remote antenna 1. Mount the transmitter following best practice mounting procedures as outlined in Section 3: Installation. 2. Connect the RF lightning arrestor to the device and tighten. IVE ATMOS OS P PL RE HE IN E X RF lightning arrestor T 154 AL EP IV E KE IG HT W H E N CI R C UI T 3.
Reference Manual Appendix C: High Gain Remote Antenna Option 00809-0100-4308, Rev BA Unwind the coaxial cable and connect the cable to both the antenna and the lightning arrestor connected to the transmitter, leaving one loop minimum for a drip loop. Ensure the drip loop is lower than the device, allowing water to flow away from the device. IVE ATMOS OS P PL IN E RE HE X 5.
Appendix C: High Gain Remote Antenna Option December 2014 156 Reference Manual 00809-0100-4308, Rev BA High Gain Remote Antenna Option
Reference Manual Appendix D: Configuration Parameters 00809-0100-4308, Rev BA December 2014 Appendix D Configuration Parameters Safety messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 157 Menu overview of the Device Descriptor (DD) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 159 Configuration parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Appendix D: Configuration Parameters December 2014 Reference Manual 00809-0100-4308, Rev BA Electrical shock can result in death or serious injury. Ground device on non-metallic tanks (e.g. fiberglass tanks) to prevent electrostatic charge build-up. Care must be taken during transportation of power module to prevent electrostatic charge build-up. Device must be installed to ensure a minimum antenna separation distance of 8 in. (20 cm) from all persons.
Reference Manual Appendix D: Configuration Parameters 00809-0100-4308, Rev BA D.2 December 2014 Menu overview of the Device Descriptor (DD) This menu is applicable for both the DD in AMS Wireless Configurator and the Field Communicator. Overview View variables and device status Configure Guided Setup Basic Setup, Verify Level, Wireless Setup, ... Manual Setup - Device Setup Wireless, Display, Units, HART, Tag, Security, Power, ...
Appendix D: Configuration Parameters Reference Manual December 2014 D.3 00809-0100-4308, Rev BA Configuration parameters This section presents a brief introduction to all configuration parameters. The Rosemount 3308 Series Transmitter can be configured for level, volume, interface level, interface distance measurements, and interface thickness. The Rosemount 3308 Series Transmitter can be pre-configured according to the ordering specifications in the Configuration Data Sheet. D.3.
Reference Manual Appendix D: Configuration Parameters 00809-0100-4308, Rev BA December 2014 Figure D-2. Upper Reference Point Adapter Upper Reference Point NPT Probe Type BSP (G) Flange Tri-Clamp™ The transmitter is designed to optimize measurement performance for each probe type. The transmitter automatically makes an initial calibration based on the type of probe that is used.
Appendix D: Configuration Parameters December 2014 Reference Manual 00809-0100-4308, Rev BA Inner Diameter, Pipe/Chamber/Nozzle Select the inner diameter for the pipe, chamber or nozzle in which the probe is mounted. Nozzle Height The distance between the Upper Reference Point (normally the lower side of the device flange) and the end of the nozzle. Note that nozzle may extend into the tank (which should be included in the height). Measurement Mode Select the Measurement Mode to use in the device.
Reference Manual Appendix D: Configuration Parameters 00809-0100-4308, Rev BA Upper Product Dielectric Constant December 2014 Enter the Upper Product Dielectric Constant (DC) as accurately as possible. This value is used for setting the automatically calculated amplitude thresholds. In addition the dielectric constant of the upper product is essential for calculating the interface level and the upper product thickness. The default value for the Upper Product Dielectric Constant is 2.
Appendix D: Configuration Parameters Reference Manual December 2014 00809-0100-4308, Rev BA Typical Interface Condition The typical interface condition in the tank. Select one of the following conditions: Typical Interface Condition Description Unknown or Other condition The typical interface condition is unknown, or varies in such a way that no typical interface condition can be stated. Layer on top (thin) The interface thickness is typically thin compared to the bottom layer.
Reference Manual Appendix D: Configuration Parameters 00809-0100-4308, Rev BA December 2014 Wireless - Broadcasting Message Content Which content (HART command) to broadcast for a message. Message Variables Which variables that are included in the content. Trigger Mode How message will be triggered. Trigger Level At which level message will be triggered. First and Trigger Variable The 1st variable contained within message which also will be used to trigger a broadcast.
Appendix D: Configuration Parameters Reference Manual December 2014 00809-0100-4308, Rev BA Units The units for length, volume, and temperature are selectable. After appropriate units have been selected, all configuration parameters and transmitter variables will be expressed in these units. Length Unit Used unit for Level and Interface Level values. Volume Unit Used unit for Volume values. Temperature Unit Used unit for Electronics Temperature value.
Reference Manual Appendix D: Configuration Parameters 00809-0100-4308, Rev BA December 2014 Security Write Protection The device configuration can be write protected. Over the Air Upgrade Wireless upgrade of radio software is possible. HART Lock Status The state of HART write lock in the device. Device Information Tag Identifier for the device (max 8 characters) used by host systems. It is recommended to enter both a short and a long tag (they may be the same).
Appendix D: Configuration Parameters Reference Manual December 2014 00809-0100-4308, Rev BA Power Mode Configures the device to take periodic measurements to conserve battery life, or to take continuous measurements. Note Always On mode is only recommended for devices connected to line power. Power Source D.3.3 Optimizes the device to make use of the power source to which it is attached. Manual Setup - Level Probe Weight Type Type of weight at the end of the probe.
Reference Manual Appendix D: Configuration Parameters 00809-0100-4308, Rev BA Upper Null Zone December 2014 Defines how close to the device's Upper Reference Point a level value is accepted. You can extend Upper Null Zone to block out disturbing echoes close to the tank top. View the Echo Curve to find out if there are disturbing echoes close to the tank top. This parameter should only be changed if there are measurement problems in the upper part of the tank.
Appendix D: Configuration Parameters Reference Manual December 2014 00809-0100-4308, Rev BA Probe Angle (Only applicable to rigid probes) Defines the angle compared to the plumb line at which the device with probe is mounted (0 means that probe is mounted vertically). Enter the angle between the probe and the vertical line. Do not change this value if the transmitter is mounted with the probe along the vertical line (which is normally the case).
Reference Manual Appendix D: Configuration Parameters 00809-0100-4308, Rev BA December 2014 Environment Vapor Dielectric Constant Enter the dielectric constant (DC) for the vapor gas in the tank. For normal air the vapor DC is close to 1. In some applications there is heavy vapor above the product surface having a significant influence on the level measurement. In such cases the vapor dielectric can be entered to compensate for this effect.
Appendix D: Configuration Parameters Reference Manual December 2014 00809-0100-4308, Rev BA Volume Calculation Method Select method for volume calculation based on tank shape or a strapping table. Strapping table requires entering level-volume pairs in a table. Diameter (L1) The diameter of the tank. Length (L2) The length (or height if the tank is shaped as a vertical cylinder) of the tank, measured between tank ends.
Reference Manual Appendix D: Configuration Parameters 00809-0100-4308, Rev BA D.3.4 December 2014 Alert Setup Signal Quality Alert Signal Quality is the product surface echo amplitude compared to the surface threshold and noise. The Signal Quality spans from 0 to 10. A low value means that there is a risk for the noise peak to be mistaken for the product surface peak. Build up on the probe and different surface conditions are factors that can result in a low Signal Quality value.
Appendix D: Configuration Parameters Reference Manual December 2014 00809-0100-4308, Rev BA High/Low Level Alerts High/Low Level Alerts are triggered when the level goes outside the user defined Limits. There are 4 standard Level Alerts. Hi Level Alert and Hi-Hi Level Alert are used for rising levels, and Lo Level Alert and Lo-Lo Level Alert are used for falling levels. See Figure D-4 and Figure D-5 for more information. Enable Level Alert If the alert will be enabled or not.
Reference Manual Appendix D: Configuration Parameters 00809-0100-4308, Rev BA December 2014 User Defined Alert Enable User Defined Alert If the alert will be enabled or not. Variable The variable selected for the alert. Alert Direction Whether the alert will be triggered above or below the variable's value. Limit The variable value that will trigger the alert. Note Alert Limit values must be outside the Upper Null Zone, the Blind zones, and areas close to the Blind Zones with reduced accuracy.
Appendix D: Configuration Parameters Reference Manual December 2014 D.3.5 00809-0100-4308, Rev BA Echo Tuning Thresholds Figure D-6. Thresholds AMPLITUDE Interface Threshold Surface Threshold Probe End Threshold Reference Threshold DISTANCE 176 Threshold Control Thresholds can be automatically calculated by device or manually set by user. This setting is valid for all thresholds on the Thresholds tab (Surface, Interface Reference, and Probe End Thresholds).
Reference Manual Appendix D: Configuration Parameters 00809-0100-4308, Rev BA December 2014 Near Zone Threshold Threshold Control The Near Zone Threshold can be automatically calculated by device or manually set by user. Threshold Threshold to filter out noise in a zone near the device. Noise below the threshold is suppressed. This threshold replaces the Surface threshold in the zone where it is applicable.
Appendix D: Configuration Parameters Reference Manual December 2014 00809-0100-4308, Rev BA Advanced Peak Detection Method Select which peak detection to use for level measurements. For furter information on when to use the peak detection method, see section “Resolving thin oil layers” on page 91. Peak Detection Method Description Peak Center Surface detected at the first amplitude peak closest to device detected above the Surface Threshold.
Reference Manual Appendix E: Mapping of Alert Messages in the HART command 48 Additional Status Field 00809-0100-4308, Rev BA December 2014 Appendix E E.1 Mapping of Alert Messages in the HART command 48 Additional Status Field Alert messages and descriptions This appendix outlines the most important alerts in the HART® command 48 Additional Status Field for 3308 Series.
Appendix E: Mapping of Alert Messages in the HART command 48 Additional Status Field Reference Manual 00809-0100-4308, Rev BA December 2014 Table E-2. Maintenance Alerts (M:) Message Additional Status(1) Description Supply Voltage Low Byte 8::Bit 4 The supply voltage is low and may affect Device Operation. Electronics Temperature Out of Limits Byte 1::Bit 2 The temperature of the electronics board has exceeded the transmitter’s operating range.
Reference Manual Index 00809-0100-4308, Rev BA A Active Advertising . . . . . . . . . . . . . . . . . . . . . . . . . . . 48, 82 Active Alerts . . . . . . . . . . . . 49, 65, 67, 72, 74, 159, 179 Agitators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7, 17 Alert Direction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175 Alert Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 159 Alerts . . . . . . . . . . . . . . . . . . . . . . . . . . .
Reference Manual Index 00809-0100-4308, Rev BA December 2014 Index L O Length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 172 Length Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 166 Limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75, 95, 173, 174 Locate Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90 Long Tag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 167 Loose flange . .
Reference Manual Index 00809-0100-4308, Rev BA Shipping Considerations. . . . . . . . . . . . . . . . . . . . . . . . . . 1 Show Level Below Probe End as Zero . . . . . . . . . . . . . . 170 Signal Quality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66 Signal Quality Alert . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173 Deadband . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75, 173 Limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Index December 2014 4 Reference Manual 00809-0100-4308, Rev BA Index
Reference Manual 00809-0100-4308, Rev BA 2014 Emerson Process Management Rosemount Inc. 8200 Market Boulevard Chanhassen, MN 55317 USA Tel (USA) 1 800 999 9307 Tel (International) +1 952 906 8888 Fax +1 952 906 8889 Emerson Process Management Blegistrasse 23 P.O. Box 1046 CH 6341 Baar Switzerland Tel +41 (0) 41 768 6111 Fax + 41 (0) 41 768 6300 Emerson FZE P.O.