Data Sheet
i2P-100 Transducer
D103197X012
Product Bulletin
62.1:i2P-100
July 2014
11
Table 6. Hazardous Area Classifications—ATEX
Certificate Certification Obtained Entity Rating Temperature Code Enclosure Rating
ATEX
Intrinsically Safe
II 1 GD
Gas
Ex ia IIC T3/T4/T5
Dust
T95°C (Tamb ≤ 85°C)
Ui = 30 VDC
Ii = 100 mA
Pi = 1.0 W
Ci = 0 nF
Li = 0 mH
T3 (Tamb ≤ 85°C)
T4 (Tamb ≤ 81°C)
T5 (Tamb ≤ 46°C)
IP66*
Flameproof
II 2 G D
Gas
Ex d IIC T5/T6
Dust
T95°C (Tamb ≤ 85°C)
- - -
T5 (Tamb ≤ 85°C)
T6 (Tamb ≤ 75°C)
IP66*
Type n
II 3 G D
Gas
Ex nC II T5/T6
Dust
T95°C (Tamb ≤ 85°C)
- - -
T5 (Tamb ≤ 85°C)
T6 (Tamb ≤ 75°C)
IP66*
*When remotely vented.
Table 7. Hazardous Area Classifications—IECEx
Certificate Certification Obtained Entity Rating Temperature Code Enclosure Rating
IECEx
Intrinsically Safe
Gas
Ex ia IIC T3/T4/T5
Ui = 30 VDC
Ii = 100 mA
Pi = 1.0 W
Ci = 0 uF
Li = 0 uH
T3 (Tamb ≤ 85°C)
T4 (Tamb ≤ 81°C)
T5 (Tamb ≤ 46°C)
IP66*
Flameproof
Gas
Ex d IIC T5/T6
- - -
T5 (Tamb ≤ 85°C)
T6 (Tamb ≤ 75°C)
IP66*
Type n
Gas
Ex nC IIC T5/T6
- - -
T5 (Tamb ≤ 85°C)
T6 (Tamb ≤ 75°C)
IP66*