Data Sheet

i2P-100 Transducer
D103197X012
Product Bulletin
62.1:i2P-100
July 2014
10
Table 4. Hazardous Area Classifications—CSA (Canada)
Certification Body Certification Obtained Entity Rating Temperature Code Enclosure Rating
CSA
Intrinsically Safe
Ex ia IIC T3/T4/T5 per drawing GE07471
Ex ia Intrinsically Safe
Class I, II, III Division 1 GP A,B,C,D,E,F,G
per drawing GE07471
Vmax = 30 VDC
Imax = 100 mA
Pi = 1.0 W
Ci = 0 uF
Li = 0 uH
T3 (Tamb ≤ 85°C)
T4 (Tamb ≤ 81°C)
T5 (Tamb ≤ 46°C)
CSA Type 4X Encl.*
Explosion Proof
Ex d IIC T5/T6
Class I, Division 1 GP, A,B,C,D T5
- - -
T5 (Tamb ≤ 85°C)
T6 (Tamb ≤ 75°C)
CSA Type 4X Encl.*
Type n
Ex nC IIC T5/T6
- - -
T5 (Tamb ≤ 85°C)
T6 (Tamb ≤ 75°C)
CSA Type 4X Encl.*
Class I, Division 2, GP A,B,C,D T5
Class II, Division 1 GP E,F,G T5
Class II, Division 2, GP F,G T5
Class III
- - - T5 (Tamb ≤ 85°C) CSA Type 4X Encl.*
*When remotely vented.
Table 5. Hazardous Area Classifications—FM (United States)
Certification Body Certification Obtained Entity Rating Temperature Code Enclosure Rating
FM
IS Intrinsically Safe
Class I, II, III Division 1 GP A,B,C,D,E,F,G
per drawing GE07470
Class I Zone 0 AEx ia IIC T3/T4/T5 per
drawing GE07470
Vmax = 30 VDC
Imax = 100 mA
Pi = 1.0 W
Ci = 0 uF
Li = 0 uH
T3 (Tamb ≤ 85°C)
T4 (Tamb ≤ 81°C)
T5 (Tamb ≤ 46°C)
NEMA 4X*
XP Explosion Proof
Class I, Division 1, GP A,B,C,D T5/T6
Class I Zone 1 AEx d IIC T5/T6
NI Non-indendive
Class I, Division 2, GP A,B,C,D T5/T6
Type n
Class I, Zone 2 AExnC IIC T5/T6
DIP Dust-igntion proof
Class II, III Division 1, GP E,F,G T5/T6
S Suitable for Use
Class II, III Division 2, GP F,G T5/T6
- - -
T5 (Tamb ≤ 85°C)
T6 (Tamb ≤ 75°C)
NEMA 4X*
*When remotely vented.