Instruction Manual
Detailed Setup
September 2013
4-21
ACTUAL TRACE FROM TEST (TYPICAL)
ACTUAL TRACE FROM TEST (TYPICAL)
ACTUAL TRACE FROM TEST (TYPICAL)
Figure 4-9. Time Series Plots of Travel Set Point, Travel, Error, and Actuator Pressure for a Typical Emergency Shutdown Valve
Pause Time—Setup Wizard sets the Partial Stroke
Pause Time to 5 seconds. This is the pause time
between the up and down strokes of the test. It can be
set for 5, 10, 15, 20 or 30 seconds.
PST Press Limit (single-acting actuators)—During
Setup Wizard or Auto Travel Calibration, the Partial
Stroke Pressure Limit will be set to a positive value for
single-acting actuators. For those actuators that vent
from the test starting point, the pressure limit will be a
minimum value. For those actuators that fill from the
test starting point, the pressure limit will be a
maximum value. The pressure signal used for this
threshold depends on relay type and is summarized
below.
Relay Type Pressure Signal
A or C Port A − Port B
B Port B − Port A
B Special App. Port B
C Special App. Port A
PST Press Limit (double-acting actuators)— During
Setup Wizard or Auto Travel Calibration, the PST
Press Limit will be set to a negative value for actuators
where the Partial Stroke Start Point is opposite of the
Zero Power Condition (e.g., Partial Stroke Start Point
= Open and Zero Power Condition = Closed) and to a
positive valve for actuators where the Partial Stroke
Start Point is the same as the Zero Power Condition.
4