Instruction Manual

Instruction Manual Supplement
D103639X012
DVC6200 Digital Valve Controller
August 2014
12
Command 48: Read Additional Status
This command was significantly revised in HART 7, but can operate in backward compatible mode for HART 5 masters
which do not supply data bytes in the request.
Supplying data bytes allows you to acknowledge the current status, and cause the more_status bit to be withdrawn
until some status bit changes value.
Exactly 0 or 9 data bytes must be supplied. If supplied, the data must exactly match the current command 48 response
in order for the more_status bit to be withdrawn.
Byte Format Description
Request
Data
bytes
0 or 9 None or nine bytes of status.
Response
Data
Bytes
0 Uint8 Command 48 Response Byte 0
1 Uint8 Command 48 Response Byte 1
2 Uint8 Command 48 Response Byte 2
3 Uint8 Command 48 Response Byte 3
4 Uint8 Command 48 Response Byte 4
5 Uint8 Command 48 Response Byte 5
6 Uint8 HART 7 Extended Field Device Status byte
7 Uint8 HART 7 Operating Mode byte
8 Uint8 HART 7 Standardized Status 1 byte
Code Class Description
Command
 specific
Response
Codes
5
Too few data
bytes
0 < x < 9, or x > 9
14 Data Mismatch Data was supplied but does not match current status