29
Volume Units—mm
3
, ml, L, in
3
Weight Units—oz, lb, g, or kg
Temperature Units—_F, _R, _C, or K
Torque Rate Units—NSm/deg, dyneScm/deg, lbfSin/deg
Fluid Density Units—degAPI, SGU (Specific Gravity) lb/in
3
, lb/ft
3
, lb/gal, degBaum hv, degBaum lt, kg/m
3
, g/cm
3
, kg/L,
g/ml, or g/L
Sensor Parameters
Enter the sensor parameters. Selections shown in the drop down are based on the sensor units chosen.
Displacer Length
Displacer Volume
Displacer Weight
Driver Rod Length
Note
Table 4‐1 provides the driver rod length of 249 sensors with vertical displacers. If your sensor isn't included in table 4‐1 refer to
figure 4‐5 to determine the driver rod length.
Instrument Mount Position
Table 4‐1. Driver Rod Length
(1)
SENSOR TYPE
(2)
DRIVER ROD
mm INCH
249 203 8.01
249B 203 8.01
249BF 203 8.01
249BP 203 8.01
249C 169 6.64
249CP 169 6.64
249K 267 10.5
249L 229 9.01
249N 267 10.5
249P (CL125-CL600) 203 8.01
249P (CL900-CL2500) 229 9.01
249V (Special)
(1)(3)
See serial card See serial card
249V (Std)
(3)
343 13.5
249VS 343 13.5
249W 203 8.01
1. Driver rod length is the perpendicular distance between the vertical centerline of the displacer and the horizontal centerline of the torque tube. See figure 4‐5. If you cannot determine the
driver rod length, contact your Emerson Process Management sales office and provide the serial number of the sensor.
2. This table applies to sensors with vertical displacers only. For sensor types not listed, or sensors with horizontal displacers, contact your Emerson Process Management sales office for the
driver rod length. For other manufacturers' sensors, see the installation instructions for that mounting.
3. The 249V is only available in Europe.