Instruction Manual
Instruction Manual
D103434X012
Configuration
November 2014
28
The Upper and Lower Sensor Limit limit what the DLC3020ft can read; values above and below these limits will not be
detected by the instrument. This is a dynamic reading based on temperature used when Temperature Compensation
is enabled.
Sensor Hardware Information
Enter the following information by selecting Sensor Hardware Information.
Model Type, End Connection Style, End Connection Type, Body Material, Pressure Rating, Mechanical Sensor Serial
Number, Displacer Size Displacer Material, Displacer Rating, G Dimension, Torque Tube Material, Torque Tube Wall,
Heat Insulator.
Sensor information is typically found on the sensor nameplate, as shown in figure 4‐4.
Note
This data is informational only and is not used in calibration or PV calculations.
Figure 4‐4. Typical Sensor Nameplate
DISPLACER
PRESSURE RATING
1500 PSI
103 CU‐IN
316 SST
249B
76543210
PSI
2 x 32 INCHES
4 3/4 LBS
K MONEL/STD
285/100 F
WCB STL
MONEL
DISPLACER
MATERIAL
DISPLACER
VOLUME
DISPLACER SIZE
(DIAMETER X LENGTH)
TORQUE TUBE MATERIAL
SENSOR TYPE
DISPLACER
WEIGHT
ASSEMBLY
PRESSURE RATING
ASSEMBLY MATERIAL
TRIM MATERIAL
23A1725‐E sht 1
E0366
Sensor Units
Select the appropriate sensor units for your application.
Note
Default units from factory are SI (Metric).
If you choose Mixed Units you must select the units for each sensor parameter.
Unit System—English Units, Metric/SI Units, Mixed Units
Length Units—mm, cm, m, in, or ft