Instruction Manual

Instruction Manual
D102748X012
Service Tools
October 2014
69
Figure 5‐1. PV % Range Indication for Direct and Reverse Action with a 32‐Inch Displacer Ranged for 8 to 24 Inches
DIRECT ACTION
REVERSE ACTION
150
130
110
90
70
50
30
10
–10
–30
–50
0 4 8 1216 20242832
LRV URV
LEVEL (INCHES)
PV (% RANGE)
150
130
110
90
70
50
30
10
–10
–30
–50
0 4 8 1216 20242832
LRVURV
LEVEL (INCHES)
PV (% RANGE)
E0383
AO— Indicates the current analog output value of the instrument, in milliamperes.
Inst Temp— Indicates the current Instrument Temperature.
Process Temperature
D Proc Temp Source— The source of measurement for Process Temperature.
D Proc Temp—- Indicates the current Process Temperature.
Torque Rate— Torque rate of the torque tube applied in PV measurements.
Upper Fluid Density — Density of Upper Fluid applied in PV measurements.
Note
Upper Fluid Density is only visible if PV is Interface
Lower Fluid Density— Density of Lower Fluid applied in PV measurements.
Note
Lower Fluid Density is only visible if PV is Level or Interface