Figure 7. Style Number of Equalizing Connections
STYLE 1
TOP AND BOTTOM CONNECTIONS,
SCREWED (S-1) OR FLANGED (F-1)
STYLE 2
TOP AND LOWER SIDE CONNECTIONS,
SCREWED (S-2) OR FLANGED (F-2)
STYLE 3
UPPER AND LOWER SIDE CONNECTIONS,
SCREWED (S-3) OR FLANGED (F-3)
STYLE 4
UPPER SIDE AND BOTTOM CONNECTIONS,
SCREWED (S-4) OR FLANGED (F-4)
28B5536-1
B1820-2
Installation
A 249 sensor may be shipped with the DLC3010 digital
level controller installed or they may be shipped
separately.
When shipping a skid mounted system, where the
displacer cannot be restrained, it is recommended that
the transmitter be uncoupled and the lever assembly
locked to prevent damage. The transmitter must be
re-coupled at commissioning, and a zero-trim will be
required.
Dimensions for the DLC3010 and 249 sensor product
construction are shown in figure 9 and tables 6 and 7.
Dimensions of other combinations are available upon
request.
Ordering Information
When ordering, specify:
n Process temperature and pressure and ambient air
temperature
n Application
n Liquid level service (specific gravity)
n Interface level service (specific gravity of both
liquids and minimum differential gap or span
required)
n Density service (minimum and maximum specific
gravity required)