Technical data
Chapter 5 Microprocessor Control 13
DataMate 3000 Series Air Conditioner Technical Manual
5.6.2 Alarm History
Select Alarm History from the Alarm Menu screen to view the Alarm History screen. The screen displays the detailed
inactive alarm information or “No alarms” if no inactive alarm exists.
The alarm No./the total number inactive, alarm name, the time/date occurrence are indicated for each alarm, as shown
in Figure 5-10.
History:
Hi Temp
01/ 01
01/ 01
1/4
00:00:00
00:00:00
Figure 5-10 Alarm History screen
Use Up/Down buttons to scroll if there are more alarms.
This system can store up to 30 latest inactive alarms. They will not be lost when the system is powered off.
5.6.3 Alarm Setpoint
The alarm setpoints will not be lost when power fails. Select Alarm Setpoint from the Alarm Menu to browse or change
the setpoints listed in Figure 5-11.
The defaults are sufficient for most applications and changing them is normally unnecessary. For special application,
they must be changed under the guidance of trained professional personnel.
Hi Temp
Lo Temp
Hi Hum
Lo Hum
29°C
29°C
60%rh
40%rh
Figure 5-11 Alarm Setpoint screen
To change the setpoint, use Up/Down buttons to move the cursor to the item desired to be changed and press Enter to
select it. Press Enter again, use Up/Down buttons to change the value of each digit, and then press Enter to validate
the change.
Refer to Table 5-1 for the defaults and the setting ranges.
5.6.4 Alarm Output
The alarm output settings will not be lost when power fails. Select Alarm Outputs from the Alarm Menu to browse or
change the settings listed in Figure 5-12.
Hi Press
Lo Press
Hi Temp
Lo Temp
ENAB
ENAB
ENAB
ENAB
Hi Hum
Lo Hum
Power Fail
ShortCycle
ENAB
ENAB
ENAB
ENAB
Custom #1
Custom #2
Rpr MnFan
Rpr Humid
ENAB
ENAB
ENAB
ENAB
Rpr Filter
Comm Fail
Coil Froze
Humid Fail
ENAB
ENAB
ENAB
ENAB
SnsrB Lost
Dschg Temp
Power Lost
Power OLV
ENAB
ENAB
ENAB
ENAB
Power PL
Power FS
ENAB
ENAB
Figure 5-12 Alarm output menu
The alarm output can be set to OPEN, CLOSE or STOP. Move the cursor to the item to be changed, press Enter to
move the cursor to the OPEN/ CLOSE /STOP field. Use Up/Down buttons to toggle among OPEN, CLOSE and STOP.
Press Enter to validate the setting.










