Specifications
Reference Manual
MHM-97408, Rev. 2
June 2008
CSI 9420
3-10
Sensor Leads
If the sensor is installed in a high-voltage environment and a fault condition or 
installation error occurs, the sensor leads and transmitter terminals could 
carry lethal voltages. Use extreme caution when making contact with the 
leads and terminals.
Use the following steps to wire the sensor and supply power to the 
transmitter:
1. Remove the transmitter enclosure cover.
2. Attach the sensor leads according to the wiring diagrams, as shown 
in Figure 3-10.
3. Connect the power module.
4. Verify the connection by observing the LCD.
5. Reattach and tighten the cover.
NOTE:
Crimp-on ferrel or lugs may be used to improve long-term reliability of sensor 
wiring.