Specifications

Reference Manual
MHM-97408, Rev. 2
June 2008
CSI 9420
3-10
Sensor Leads
If the sensor is installed in a high-voltage environment and a fault condition or
installation error occurs, the sensor leads and transmitter terminals could
carry lethal voltages. Use extreme caution when making contact with the
leads and terminals.
Use the following steps to wire the sensor and supply power to the
transmitter:
1. Remove the transmitter enclosure cover.
2. Attach the sensor leads according to the wiring diagrams, as shown
in Figure 3-10.
3. Connect the power module.
4. Verify the connection by observing the LCD.
5. Reattach and tighten the cover.
NOTE:
Crimp-on ferrel or lugs may be used to improve long-term reliability of sensor
wiring.