Product data

www.assetweb.com8
Machinery Health
Management October 2014
Physical Specifications
Power Options
SmartPower
Module Intrinsically safe
Replaceable
Lithium-Thionyl Chloride
External DC-power Input Voltage: 10 - 28 VDC
Provides unlimited operating life — recommended for applications requiring faster
update rates
Operating Life
1
Values apply to Device Software version 5.020 and higher
Conguration 1 2 3
Overall Values (min) 30 60 240
2
Energy Bands (hr) 8 8 8
Spectrum (hr) 24 24 24
High Resolution Waveform & Spectrum (days) 30 30 30
Operating life (years) 1 to 2 1.5 to 3 2 to 4
Materials of Construction
Enclosure Housing — Low-copper aluminum (standard)
Paint — Polyurethane
Cover O-ring — Buna-N
Terminal Block and Battery Pack PBT
Antenna PBT/PC integrated omni-directional antenna (standard)
Optional extended range antenna available in some markets.
Mounting Vibration sensor must be mounted directly on asset being monitored
Transmitter may be mounted up to 100’ (30.5 m) away from sensor
Weight 2 kg (4.6 lbs.) without LCD
2.1 kg (4.7 lbs.) with LCD
Enclosure Ratings NEMA 4X / IP66
Performance Specifications
ElectroMagnetic Compatibility (EMC) Meets all relevant requirements of EN 61326.
Measurement Accuracy RMS Velocity
3
:
+/- 5% from 10 Hz to 800 Hz
+/-3 dB from 2 Hz to 1000 Hz
PeakVue Impacting
+/- 5% from 2000 Hz to 10 kHz
+/- 3 dB from 1000 Hz to 20 Khz
Temperature: +/- 2°C
Measurement Precision Vibration: +/- 0.2 dB
Temperature: +/- 2°C
Self Calibration The analog-to-digital measurement circuitry automatically self-calibrates for each
update by comparing the dynamic measurement to internal reference elements
(1) Operating life with a single SmartPower Module is highly dependent on user conguration. Values shown are for operation at an ambient temperature of 21°C (70°F) as part of a
well-formed wireless network. Daily measurements are thumbnail spectra. Continuous exposure to extreme temperatures may signicantly reduce operating life. Older revisions
exhibit a maximum power module life of 2 years at room temperature with a 54 minute burst rate. Actual results depend on the installation process.
(2) Burst rates over 60 minutes are congured using the "PowerSave" mode.
(3) Measurement accuracy is the absolute accuracy of the measurement relative to a known, calibrated excitation for WirelessHART devices. Values shown represent the expected
performance operating under steady-state conditions (20°C with no external interference).