Instruction manual

Model 3200 HP
High Purity Water pH Sensor
with pHaser
TM
Flowing Junction
Instruction Manual
51A-3200HP/rev.F
January 2011
SPECIFICATIONS
Minimum Conductivity: 0.4 µS/cm
Sample Temperature: 0°C to 70°C (32°F to 160°F)
Sample Inlet Pressure: 5 to 10 psig (134-170 kPa abs);
no back pressure allowed; sample must drain to open
atmosphere
Wetted Materials: Silicone, polycarbonate/ polyester,
glass, stainless steel, PVDF, Viton
1
1. Viton is a registered trademark of DuPont Performance Elastomers.
For additional information,
please visit our website at
www.emersonprocess.com/raihome/liquid/
CAUTION
INSTALLATION
1. Refer to Figure 1 for dimensions of the mounting plate and the positions and
diameter of the mounting holes.
2. Remove the bag containing the diffuser plate from the flow cell. Place the dif-
fuser plate flat in the bottom of the flow cell. See Figure 2. Do not install the
sensor yet.
3. Connect the sample and drain lines to the flow cell. Process connections
are ¼ inch FNPT.
The flow cell must drain immediately to open atmosphere. Do not use
a long or convoluted drain line.
FIGURE 1
inch
mm
PN 34083-00
Diffuser
PN 3200HP
pH sensor
Flow cell
Flow cell nut
FIGURE 2
The wetted sensor materials may not be compatible
with process com position and operating conditions.
Application compat ibility is entirely the
responsi-
bility of the user.
SENSOR/PROCESS APPLICATION COMPATIBILITY
ATEX DIRECTIVE
Special Conditions for safe use
1. All pH/ORP sensors have a plastic enclosure which must only be cleaned with
a damp cloth to avoid the danger due to a build up of an electrostatic charge.
2. All pH/ORP sensor Models are intended to be in contact with the process fluid
and may not meet the 500V r.m.s. a.c. test to earth. This must be taken into
consideration at installation.