User's Manual

Instruction Manual
748332-F
April 2003
2-6 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 NDIR
.6
[15]
1.6
[40]
20.0
[508.0]
STANDARD
24.8
[628.7]
EXTENDED
17.41
[142.2]
STANDARD
22.41
[569.2]
EXTENDED
1.0
[25]
1.1
[27]
8.4
[
213
]
1.1
[
28
]
5.6
[143]
2.8
[71]
4.3
[109]
.5
[13]
1.1
[28]
6.0
[152]
.6
[15]
8.2
[208]
8.4
[213]
6.2
[157]
G. POWER CABLE TO NETWORK.
F. NETWORK CABLE CONNECTIONS TO PLATFORM.
E. PURGE GAS IN: 1/4" O.D. TUBE FITTING.
D. REFERENCE IN: 1/4" O.D. TUBE FITTING.
C. REFERENCE OUT: 1/4" O.D. TUBE FITTING.
B. SAMPLE OUT: 1/4" O.D. TUBE FITTING.
A. SAMPLE IN: 1/4" O.D. TUBE FITTING.
5. MODULE TO BE INSTALLED WITHIN ±15° OF HORIZONTAL.
4. POWER REQUIREMENTS: 24 VDC 3.5 A.
3. ELECTRICAL INSTALLATION MUST BE IN COMPLIANCE WITH NATIONAL ELECTRICAL
CODE (ANSI/NFPA 70) AND/OR ANY APPLICABLE NATIONAL OR LOCAL CODES.
2. MODULE IS NOT WEATHERPROOF.
1. APPROXIMATE WEIGHT: 24.2 LB (11.0 kg).
DIMENSIONS
INCH
[mm]
Figure 2-5. Outline and Mounting Dimensions