Instruction Manual 748332-F April 2003 Model NGA2000 NDIR Non-Dispersive Infrared Analyzer http://www.processanalytic.
ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications.
Instruction Manual Model NGA2000 NDIR 748332-F April 2003 TABLE OF CONTENTS PREFACE...........................................................................................................................................P-1 Definitions ...........................................................................................................................................P-1 Safety Summary ................................................................................................................
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR 4-0 MAINTENANCE AND TROUBLESHOOTING .........................................................................................4-1 4-1 Overview................................................................................................................................4-1 4-2 PCB Replacement .................................................................................................................4-1 4-3 Power Fuse Replacement ...............
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR LIST OF ILLUSTRATIONS Figure 1-1. Figure 1-2. Figure 2-1. Figure 2-2. Figure 2-3. Figure 2-4. Figure 2-5. Figure 3-1. Figure 3-2. Figure 3-3. Figure 3-4. Figure 3-5. Figure 3-6. Figure 3-7. Figure 3-8. Figure 3-9. Figure 3-10. Figure 3-11. Figure 3-12. Figure 3-13. Figure 4-1. Figure 4-2. Figure 4-3. Figure 4-4. Figure 4-5. Figure 4-6. Figure 4-7. Figure 4-8. NDIR Technology ..........................................................................
Instruction Manual 748332-F April 2003 iv Contents Model NGA2000 NDIR Rosemount Analytical Inc.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR PREFACE The purpose of this manual is to provide information concerning the components, functions, installation and maintenance of the NGA2000 NDIR and the System Accessories of the NGA2000 System. Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this module before operating it. Read this instruction manual completely.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR SAFETY SUMMARY If this equipment is used in a manner not specified in these instructions, protective systems may be impaired. AUTHORIZED PERSONNEL To avoid explosion, loss of life, personal injury and damage to this equipment and on-site property, all personnel authorized to install, operate and service the this equipment should be thoroughly familiar with and strictly follow the instructions in this manual. SAVE THESE INSTRUCTIONS. DANGER.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR CAUTION HAND INJURY HAZARD Dropping the front panel of the Platform while hand or fingers are inside either case handle can cause serious injury. CAUTION PARTS INTEGRITY Tampering with or unauthorized substitution of components may adversely affect safety of this product. Use only factory approved components for repair.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH PRESSURE GAS CYLINDERS Edited from selected paragraphs of the Compressed Gas Association's "Handbook of Compressed Gases" published in 1981 Compressed Gas Association 1235 Jefferson Davis Highway Arlington, Virginia 22202 Used by Permission 1. Never drop cylinders or permit them to strike each other violently. 2.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR DOCUMENTATION The following NGA2000 NDIR instruction materials are available. the local representative to order.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR GLOSSARY OF TERMS Analyzer Module The module that contains all sensor/detector components for development of a Primary Variable signal; includes all signal conditioning and temperature control circuitry. Backplane The interconnect circuit board which the Controller Board, Power Supply, Analyzer Module power and network cables, I/O Modules and Expansion Modules plug into. Control Module The Operator Interface plus the Controller Board.
Instruction Manual Model NGA2000 NDIR 748332-F April 2003 Softkeys The five function softkeys located below the front panel display; they assume the function displayed directly above each on the display, a function dictated by software. System Any collection of Analyzer Module(s), Platform(s), I/O Module(s) and Expansion Module(s). Rosemount Analytical Inc.
Instruction Manual 748332-F April 2003 P-8 Preface Model NGA2000 NDIR Rosemount Analytical Inc.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR SECTION 1 DESCRIPTION AND SPECIFICATIONS 1-1 OVERVIEW This manual describes the Non-Dispersive Infrared (NDIR) Analyzer Module of Rosemount Analytical's NGA2000 Series of gas analysis components.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR During analysis, a portion of the infrared radiation is absorbed by the component of interest in the sample. The quantity of infrared radiation that is absorbed is proportional to the component concentration. The detector is a "gas microphone" based on the Luft principle. The detector is generally filled with the same gas being analyzed.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR 1-4 1-5 SAMPLE REQUIREMENTS Maximum allowable sample pressure is 690 hPa-gauge (10 psig) for a standard configuration NDIR that has a flow restrictor which sets the flow at between 0.5 L/min. to 1 L/min. Special high pressure cells (up to 10,350 hPa-gauge, 150 psig) are available. Sample temperature range is 0°C to 55°C, and maximum dewpoint is 40°C. The sample must be filtered to exclude particulates larger than 2 microns in size.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR SAMPLE OUT (BOTTOM) REFERENCE OUT REFERENCE IN PURGE GAS IN FAN SAMPLE IN POWER SUPPLY BOARD CHOPPER MOTOR SOURCE CASE TEMPERATURE SENSOR PRESSURE COMPENSATION BOARD (OPTION) CELLS FLOW SENSOR MICRO BOARD DETECTOR THERMAL FUSE DETECTOR TEMPERATURE CONTROL RTD DETECTOR COVER SIGNAL BOARD OSCILLATOR BOARD SHUTTER ADJUST ACCESS HOLES Figure 1-2.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR 1-7 SPECIFICATIONS a. General Measurement Species:.................. Heteroatomic gases such as ammonia (NH3), carbon dioxide (CO2), carbon monoxide (CO), carbon monoxide + carbon dioxide ethylene (C2H4), hexane (C6H14), methane (CH4), nitric oxide (NO) and sulfur dioxide (SO2) Ranges: .........................................
Instruction Manual 748332-F April 2003 1-6 Description and Specifications Model NGA2000 NDIR Rosemount Analytical Inc.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR SECTION 2 INSTALLATION WARNING. CAUTION. Before starting to install this equipment, read the “Essential Instructions” on the inside cover and the Safety Summary beginning on page P-2. Failure to follow the safety instructions could result in serious injury or death. 2-1 UNPACKING If the NDIR Analyzer Module is received as a separate unit, carefully examine the shipping carton and contents for signs of damage.
Instruction Manual 748332-F April 2003 2-3 Model NGA2000 NDIR negated if the same flow rate is used for sample, zero and span gases. But, if flow is high enough to cause elevated pressure, careful control (tighter tolerance) of flow rate is required to avoid errors. LOCATION Install the NDIR Analyzer Module in a clean, non-hazardous, weather protected, vibration free location free from extreme temperature variations.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR CELL LENGTH CELL VOLUME CC TOTAL VOLUME CC TIME FOR 2 VOLUMES @ 2 SCFH (1L/MIN) mm inch without inlet tube cell with inlet tube at 750 mm Hg 3 0.118 0.85 12 2 sec. 4 0.157 1.14 12 2 sec. 8 0.315 2.28 13 2 sec. 16 0.630 3.56 16 2 sec. 32 1.25 9.12 20 2 sec. 64 2.52 18.24 25 3 sec. 128 4.03 35.48 44 3 sec. 232 9.13 65.12 73 6 sec. 343 13.50 97.76 105 13 sec. 381 15.00 108.60 116 14 sec.
Instruction Manual 748332-F April 2003 2-6 Model NGA2000 NDIR ELECTRICAL CONNECTIONS NOTE Electrical connections must be in compliance with National Electrical Code (ANSI/NFPA 70) and/or any applicable national or electrical codes. Two electrical connections are required on the Analyzer Module; POWER and NETWORK. See Figure 2-3 below.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR SAMPLE SOURCE REFERENCE SOURCE J2-1 J2-2 J2-3 J2-4 J9 J2 J4 J5 J6 J4 POWER SUPPLY BOARD PRESSURE COMPENSATION BOARD J11 J1 SIGNAL BOARD J4 COMPUTER BOARD RECORDER OUTPUT J5 J7 J1 J3 J2 J6 J15 J14 J6 J10 J7 1 2 3 4 1 J5 2 J7 J8 J3 J2 J3 J1 CASE TEMPERATURE SENSOR °C FLOW SENSOR FAN ASSEMBLY RESISTOR MODULATION CHECK FUSE, THERMAL CUTOFF HEATSINK/SOURCE DRIVER RTD, TEMP CTRL HEATER SENSOR °C DETECTOR BASE MOTOR A
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR 8.4 [213] 8.2 [208] 6.2 [157] .6 [15] .5 [13] 2.8 [71] 4.3 [109] 1.1 [28] 20.0 [508.0] STANDARD 6.0 [152] 24.8 [628.7] EXTENDED 17.41 [142.2] STANDARD 22.41 [569.2] EXTENDED 1.6 [40] .6 [15] 8.4 [213] 1.1 [28] 1.1 [27] G. F. E. D. C. B. A. POWER CABLE TO NETWORK. NETWORK CABLE CONNECTIONS TO PLATFORM. PURGE GAS IN: 1/4" O.D. TUBE FITTING. REFERENCE IN: 1/4" O.D. TUBE FITTING. REFERENCE OUT: 1/4" O.D. TUBE FITTING.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR SECTION 3 STARTUP AND OPERATION 3-1 b. OVERVIEW Prior to initial startup, the user should leak test the module as outlined in Section 2. For the remainder of this section, Analyzer Module interconnection with a Platform or some interfacing component will be assumed. Display and Keypad information shall refer to that which the user can expect to see and do with regard to the Front Panel of the Platform.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR Diagnostic Service menus. If both settings are within ±5% tolerance of factory setting, go to Section 3-4 below for binding and Section 3-5 on page 3-4 for calibration. If not, refer first to Section 4-11 on page 4-6 for instructions about oscillator tune/source balance shutter adjustments. Controller Board and the network "self-install" (bind together) during initial startup.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR 23.2 % CO Analyzer XXXXXXXX Basic Controls Measurement range Numbers: Range upper limit: Range and functional control: Calibration… 25% Local Status: Ready Home Escape F1 F2 Zero F3 Figure 3-3. Span Info F4 F5 Basic Controls Menu 23.2 % CO Analyzer XXXXXXXX Expert controls and setup Expert analyzer controls ... Auxiliary module controls ... System set up ... Analyzer module set up ... Auxiliary module set up ...
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR 23.2 % CO Analyzer XXXXXXXX Main Menu Help The Main Menu for the analyzer system. Note that this menu refers to the particular analyzer selected from the run screen, when used in a system. The softkey marked “HOME” will always return you to this screen. Help menu system... Help on help... Keyboard controls... Editing controls... Home Escape F1 F2 Figure 3-6. 3-5 CALIBRATION Calibration can be executed from the Basic Controls menu.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR 5. Refer to the Data Sheet in the rear of this manual for Span Factors as originally set at the factory. Select Span Factor, press ENTER, adjust the value to match the values on the Data Sheet with the ↑ and ↓ arrow keys, and press ENTER. If Data Sheet is not available, enter 0.000015 with the ↑ and ↓ arrow keys, and press ENTER. Do not press ESCAPE unless retention of prior settings is desired. Analyzer Module Set Up, and Calibration Parameters.
Instruction Manual 748332-F April 2003 3-6 Model NGA2000 NDIR LINEARIZATION The NDIR Analyzer Module can be operated in linear and non-linear mode. Linearization can be toggled ON/OFF in the Expert Controls menu. In the OFF position, linearization is disabled for all ranges, and the component of interest is measured in percent of fullscale. In the ON position, measurement is in engineering units: Either ppm or percent of concentration.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR 100 % READING % FULLSCALE 50 % 0 % 0 500 1000 1500 2500 3000 3500 4000 4500 5000 CONCENTRATION ppm Display ppm Figure 3-7. 2000 Typical Linearization Curve, Linearizer OFF 1 0.9 0.8 0.7 0.6 CONCENTRATION NORMALIZED 0.5 0.4 0.3 0.2 0.1 0 0 1 Figure 3-8. Rosemount Analytical Inc. 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.
Instruction Manual 748332-F April 2003 3-7 ROUTINE OPERATION Set the NDIR Analyzer Module for desired operating range. Zero and span the module, and then supply sample gas to the SAMPLE INLET at the rear of the module. The NDIR Analyzer Module will now automatically and continuously analyze the sample stream. As a check of instrument performance, the operator should keep a log of zero/span status. Maximum permissible interval between calibrations depends on the analytical accuracy required.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR 4 Current measurement parameters Main Menu Basic controls... Expert controls and setup... (Operational configuration) Technical level configuration...
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR Expert controls and setup From MAIN MENU Expert analyzer controls... Auxiliary module controls... System set up... Analyzer Module set up... Auxiliary Module set up... Local I/O set up... ⇒ Auxiliary module controls Expert controls Measurement range number: Range lower limit: Range upper limit: Linearizer: Range and functional control: Zero/Span calibration... Ranges with valid calibration: Physical measurements...
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR From MAIN MENU ⇓ #5 Linearity coefficients Calibration gas list Expert controls and setup Expert analyzer controls... Auxiliary module controls... System set up... Analyzer Module set up... Auxiliary Module set up... Local I/O set up... Zero gas - range 1: Span gas - range 1: Zero gas - range 2: Span gas - range 2: Zero gas - range 3: Span gas - range 3: Zero gas - range 4: Span gas - range 4: Calibration... 0 ppm 9.3 ppm 0 ppm 19.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR From MAIN MENU ⇓ Select IO modules Technical configuration menu System set up... Service menus... Diagnostic menus... Other module diagnostic menus... Listing of all modules... Select the modules you wish to bind to the current analyzer. Module Tag Module Tag Module Tag Module Tag Module Tag Module Tag Module Tag Module Tag Main display configuration, Front panel control, and Date and time displays can be found in the Platform manual.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR From MAIN MENU ⇓ Technical configuration menu System set up... Service menus... Diagnostic menus... Other module diagnostic menus... Listing of all modules... Analyzer diagnostics For Service set up and Service menus, see preceding flow charts. Control Unit Diagnostics See Platform manual. Diagnostic menus Control module diagnostics... Analyzer module diagnostics...
Instruction Manual 748332-F April 2003 3-14 Startup and Operation Model NGA2000 NDIR Rosemount Analytical Inc.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR SECTION 4 MAINTENANCE AND TROUBLESHOOTING balance adjustments. These are factory-set, and should be adjusted only in the following cases: DANGER ELECTRICAL SHOCK HAZARD Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel. 4-1 if the operator has changed sources, oscillator board or detector if the operator disassembles the bench to any degree.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR TOP VIEW POWER SUPPLY BOARD PRESSURE COMPENSATION BOARD MICRO BOARD SIGNAL BOARD SIDE VIEW Figure 4-1. 4-3 Printed Circuit Board Fold-Out Panel Views shown in Figure 4-2 below, and push and turn the fuseholder cap 1/4 turn counterclockwise. Remove and replace the fuse as required. POWER FUSE REPLACEMENT Remove power to the Analyzer Module prior to fuse replacement.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR 4-4 MODULE FAN REPLACEMENT The Analyzer Module fan assembly is disassembled as shown in Figure 4-3 below. Before doing so, the entire optical bench must be removed. See Figure 4-5on page 4-5. AIR FLOW Figure 4-3. Rosemount Analytical Inc.
Instruction Manual 748332-F April 2003 4-5 Model NGA2000 NDIR CHOPPER MOTOR REPLACEMENT See Figure 4-4 below for a exploded view of the Motor/Source Assembly. The entire optical bench must be removed to perform this replacement. 4-6 SOURCE REPLACEMENT See Figure 4-4 below for a exploded view of the Motor/Source Assembly. The entire optical bench must be removed to perform this replacement.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR 4-7 DETECTOR REMOVAL See Figure 4-5 below and Figure 4-8 on page 4-9 for breakdowns of the optical bench. The entire optical bench must be removed in order to perform this disassembly. Oscillator Tune Adjust Detector Assembly Motor/Source Assembly Power Transistor/Heatsink Flow Sensor Figure 4-5. Cell, PCB Assembly – Exploded View Rosemount Analytical Inc.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR Open Analyzer Module front panel 4-8 4-9 FLOW SENSOR REPLACEMENT See Figure 4-5 on page 4-5 for Flow Sensor location. To replace Flow Sensor, remove all connecting hardware and undo connections to the sample line. CASE TEMPERATURE SENSOR REPLACEMENT Case Temperature Sensor is attached to the motor source assembly. To replace this sensor, cut the Ty-Rap binder and disconnect the sensor connector.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR DETECTOR ASSEMBLY COVER (Front View) Oscillator Tune Adjustment SHUTTER Sample OPEN CLOSE Source Balance Shutter Adjustment Screw Hole Figure 4-6. Oscillator Tune, Source Balance Shutter Adjustments Detector Oscillator Board Resistive Temperature Detector Thermal Fuse Detector Heater Figure 4-7. Rosemount Analytical Inc.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR Inspect the cell inside by holding it up to a bright light. If particles are seen, repeat Steps 5 and 6 as often as necessary. 4-12 CLEANING CELLS When zero gas is introduced in the SAMPLE IN inlet and the Detector signal is greater 1.2V, the cells and windows require cleaning. To do this, follow this procedure: After cleaning cell, examine O-rings at the source, detector and end caps. If damaged, replace with new O-rings.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR Detector Assembly SHORT PATH CELLS 4mm TO 2.5 INCHES Desiccant Holder Shutter Block Assembly Spacer Sample Cell Desiccant Holders Motor/Source Assembly Detector Assembly LONG PATH CELLS Sample Cell Shutter Block Assembly Motor/Source Assembly Figure 4-8. Rosemount Analytical Inc.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR outlets to keep moisture from entering the reference cell (see Figure 4-7 on page 4-7). The desiccant should be replaced each time the cell is opened, refer to Table 4-1 below. Before inserting desiccant, purge cell with dry inert gas (i.e., N2). 4-13 CELL DESICCANT The reference cell may use a flowing reference. If so, desiccant is required.
Instruction Manual Model NGA2000 NDIR 4-14 MODULATION CHECK The modulation check is a measure of the energy absorption capacity of a gas mixture, where total absorbency is 100% modulation and zero absorbency is 0% modulation. The modulation check can be used as a tool for tracking the sensitivity of the Analyzer Module and for discovering potential detector/source problems. To perform a modulation check, do the following: Access the Primary Variable Parameters menu. Select Modulation Check.
Instruction Manual 748332-F April 2003 4-12 Maintenance and Troubleshooting Model NGA2000 NDIR Rosemount Analytical Inc.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR SECTION 5 REPLACEMENT PARTS WARNING PARTS INTEGRITY Tampering with or unauthorized substitution of components may adversely affect safety of this product. Use only factory-approved components for repair. 5-1 MATRIX Each analyzer is configured per the customer sales order. Below is the NDIR sales matrix which lists the various configurations available. NDIR To identify the configuration of an analyzer, locate the analyzer name-rating plate.
Instruction Manual 748332-F April 2003 5-2 REPLACEMENT PARTS While the following sections list parts which are common to all NDIR Analyzer Modules, the configuration number is required when ordering parts which are specific to an individual application. The configuration number is on the Data Sheet in the rear of this manual.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR SECTION 6 RETURN OF MATERIAL 6-1 RETURN OF MATERIAL If factory repair of defective equipment is required, proceed as follows: 1. Secure a return authorization from a Rosemount Analytical Inc. Sales Office or Representative before returning the equipment. Equipment must be returned with complete identification in accordance with Rosemount instructions or it will not be accepted. Rosemount CSC will provide the shipping address for your instrument.
Instruction Manual 748332-F April 2003 6-2 Return of Material Model NGA2000 NDIR Rosemount Analytical Inc.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR SECTION 7 INDEX A Analyzer Module applying power, 3-1 calibration procedure, 3-4 function, 1-1 installation into Platform, 2-1 installation location, 2-2 more that one in system, 3-1 operating temperature, 2-2 unable to calibrate, 3-4 linear mode, 3-6 linearization coefficients calculating, 3-6 linearizer, 1-1 Luft principle, 1-2 M Maximum permissible interval between calibrations, 3-8 maximum range changing, 3-8 B N binding, 3-2 non-linear m
Instruction Manual 748332-F April 2003 7-2 Index Model NGA2000 NDIR Rosemount Analytical Inc.
WARRANTY Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in workmanship and material under normal use and service for a period of twelve (12) months from the date of shipment by Seller. Consumables, glass electrodes, membranes, liquid junctions, electrolyte, o-rings, etc., are warranted to be free from defects in workmanship and material under normal use and service for a period of ninety (90) days from date of shipment by Seller.
Instruction Manual 748332-F April 2003 Model NGA2000 NDIR EMERSON PROCESS MANAGEMENT Rosemount Analytical Inc. Process Analytic Division 1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 E gas.csc@emersonprocess.com EUROPEAN TECHNOLOGY CENTER Fisher-Rosemount GmbH & Co. Industriestrasse 1 63594 Hasselroth Germany T 49-6055-884 0 F 49-6055-884209 ASIA - PACIFIC Fisher-Rosemount Singapore Private Ltd.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 29 LINRANGE4 Menu: 30 LINRANGE0 Menu: 31 AMPWRI1 Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 32 FLOCHEK1I1 Menu: 33 FILTER Menu: 34 AM1VI1 7-12 Return of Material Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 35 AMTEMPI1 Menu: 36 AM2VA Menu: 37 PLIMITSA Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 38 TLIMITSA Menu: 39 AMMISCI1 Menu: 40 MODTEST 7-14 Return of Material Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 41 ANALSIMPLE Menu: 42 FILTERI1 Menu: 43 LINRANGE0I1 Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 44 CALFACTORS Menu: 45 R1FACTORS Menu: 46 RN2FACTORS 7-16 Return of Material Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 47 RN3FACTORS Menu: 48 RN4FACTORS Menu: 49 AMHELPINDEX Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 50 LINRANGE1I1 Menu: 51 AMTRENDI1 Menu: 52 CALFACTORSI1 7-18 Return of Material Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 53 APARLST2 Menu: 54 APARLST4 Menu: 55 APARLST5 Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 56 APARLST6 Menu: 57 LISTNOTES Menu: 58 MPARMS2 7-20 Return of Material Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 59 DISPLAY Menu: 60 ANALSET2 Menu: 61 ESCAPE_HELP Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 62 MPARMSI1 Menu: 63 TLIMITSIAI1 Menu: 64 BAROM_PARMS 7-22 Return of Material Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 65 ANALSETI2 Menu: 66 AMHELPINDEX2 Menu: 67 MODTESTI1 Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 68 BAROM_PARMI1 Menu: 69 SW_DIAG Menu: 70 CALI1 7-24 Return of Material Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 71 AM1VHIST Menu: 72 LINFUNCT Menu: 73 POLYSETUP Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 74 MIDPOINT1 Menu: 75 MIDPOINT2 Menu: 76 MIDPOINT3 7-26 Return of Material Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 77 MIDPOINT4 Menu: 78 POLYGAS1 Menu: 79 POLYGAS2 Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 80 POLYGAS3 Menu: 81 POLYGAS4 Menu: 82 POLYGAS5 7-28 Return of Material Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 83 POLYGAS6 Menu: 84 POLYGAS7 Menu: 85 POLYGAS8 Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 86 POLYGAS9 Menu: 87 POLYGAS0 Menu: 88 EXP_CAL 7-30 Return of Material Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 89 ZEROI2 Menu: 90 EXP_CAL_DAT Menu: 91 EXP_CAL_DATI Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 92 UNITS Menu: 93 UNITSI1 Menu: 94 RESET 7-32 Return of Material Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 95 STORE Menu: 96 RFHIST1A Menu: 97 RFACTORSIA Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 98 RFHIST2A Menu: 99 RFHIST3A Menu: 100 RFHIST4A 7-34 Return of Material Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 101 TWEAKI1 Menu: 102 INIT Menu: 103 SW_DIAGI1 Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 104 STOREDPVA Menu: 105 ZERO_NOW2 Menu: 106 SPAN_NOW2 7-36 Return of Material Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 107 POLYSETI1 Menu: 108 POLYSETI2 Menu: 109 POLYSETI3 Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 110 ANALSETI3 Menu: 111 MPARMS Menu: 112 FLOWLIMITS 7-38 Return of Material Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 113 PLIMITSAI1 Menu: 114 ABOUT Menu: 115 ABOUT1 Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR Menu: 116 ALARM1 7-40 Return of Material Rosemount Analytical Inc.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR SECTION 8 INDEX A Analyzer Module applying power, 3-1 calibration procedure, 3-4 function, 1-1 installation into Platform, 2-1 installation location, 2-2 more that one in system, 3-1 operating temperature, 2-2 unable to calibrate, 3-4 linear mode, 3-6 linearization coefficients calculating, 3-6 linearizer, 1-1 Luft principle, 1-2 M Maximum permissible interval between calibrations, 3-8 maximum range changing, 3-8 B N binding, 3-2 non-linear m
Instruction Manual 748417-B April 2003 8-2 Index Model NGA2000 NDIR Rosemount Analytical Inc.
Addendum 748461-B April 2003 Model NGA2000 PMD, NDIR VAISALA BAROMETRIC PRESSURE TRANSDUCER vidual transducer at a known ambient air pressure. Test points are provided on the circuit board for this adjustment. This addendum serves as an amendment to the Model NGA 2000 PMD and NDIR instruction manuals.
Addendum 748461-B April 2003 Model NGA2000 PMD, NDIR Barometric Pressure menu reads the correct ambient air pressure. 5. Connect the instrument sample inlet tubing to the Vaisala Pressure Inlet port. 6. Enable the Barometric Pressure Compensation and the Use Local Reading options via the Expert Controls – Analyzer Module Diagnostics – Barometric Pressure Compensation menus of the instrument. The instrument will automatically detect and report the ‘Presence’ of the Pressure Compensation Option.
WARRANTY Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in workmanship and material under normal use and service for a period of twelve (12) months from the date of shipment by Seller. Consumables, glass electrodes, membranes, liquid junctions, electrolyte, o-rings, etc., are warranted to be free from defects in workmanship and material under normal use and service for a period of ninety (90) days from date of shipment by Seller.
Instruction Manual 748417-B April 2003 Model NGA2000 NDIR EMERSON PROCESS MANAGEMENT Rosemount Analytical Inc. Process Analytic Division 1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 E gas.csc@emersonprocess.com EUROPEAN TECHNOLOGY CENTER Fisher-Rosemount GmbH & Co. Industriestrasse 1 63594 Hasselroth Germany T 49-6055-884 0 F 49-6055-884209 ASIA - PACIFIC Fisher-Rosemount Singapore Private Ltd.