User's Manual
5-64 APPLICATION FUNCTIONS
MON2000
FID Configuration JULY 2010
(b) Click Modify. The blue background of
the parameter fields will turn white,
indicating that they are now modifiable.
The parameters that you can configure
are:
• FID ignition - Select Manual if you wish
to control the ignition of the FID
(default); select Auto to let the GC
control the ignition of the FID.
• Number of ignition attempts - Indicates
the number of times the GC will try to
light the flame. If an 'Auto' FID ignition
sequence fails to light the flame after the
specified number of attempts, the GC
will close the hydrogen valve, switch the
FID ignition parameter to Manual, and
set an active alarm.
• Wait time between tries.
• Igniter on duration.
• Flame On sense temperature.
• Flame Out sense temperature.
• Electrometer gain - Your choices are
High and Low (default).
(c) Click Save to accept the changes or
Cancel to reject them.