Specifications

Configuration and Use Manual 29
Calibration
Startup ConfigurationCalibrationBefore You Begin
Table 3-2 PROFIBUS parameters for Smart Meter Verification
Step number Step description Parameters
1 Set output state Block: Transducer block 1
Index: 182
Value:
0: Last measured value (default)
1: Fault
2 Start/abort procedure Block: Transducer block 1
Index: 72 (Start/Stop Meter Verification)
•0x00: No effect
0x01: Start On-Line Meter Verification
3 Check current algorithm state Block: Transducer block 1
Index: 75
Value:
Bits 4–6: State
4 Read percent complete Block: Transducer block 1
Index: 189 (Progress)
5 Check algorithm abort state Block: Transducer block 1
Index: 75
Value:
Bits 0–3: Abort code
6 Check inlet stiffness Block: Transducer block 1
Index: 77
0: Within uncertainty limit
1: Outside uncertainty limit
7 Check outlet stiffness Block: Transducer block 1
Index: 78
0: Within uncertainty limit
1: Outside uncertainty limit
8 Read abort code Block: Transducer block 1
Index: 185
Codes: See Table 3-3