Specifications

142 Model 2700 Transmitter with PROFIBUS-PA
Slave Diagnostic Response Bytes
Table E-2 Byte 2
Bit Indication
0 Slave must be parameterized
1 Static diagnostic: master requesting diagnostics until bit is reset
2 This bit is always set to 1
3 Response monitoring/watchdog (1 = ON; 0 = OFF)
4 Slave is in freeze mode (1 = ON; 0 = OFF)
5 Slave is in sync mode (1 = ON; 0 = OFF)
6 Reserved
7 Slave is deactivated in master parameter set (this is set by the master)
Table E-3 Byte 3
Bit Indication
0 Reserved (this is always set to 0)
1 Reserved (this is always set to 0)
2 Reserved (this is always set to 0)
3 Reserved (this is always set to 0)
4 Reserved (this is always set to 0)
5 Reserved (this is always set to 0)
6 Reserved (this is always set to 0)
7 Diagnostic overflow—transmitter has more diagnostic data than it can report
Table E-4 Byte 4
Bit Indication
0
Master station address
An address in the range of 0–125 decimal (0x0–0x7D hex) is the address of the
controlling master.
An address of 255 decimal (0xFF hex) means the slave is not controlled or
parameterized by a master.
1
2
3
4
5
6
7