Pressure Compensation and Temperature Compensation
There are two ways to configure external temperature compensation:
If the operating temperature is a known static value, you can configure that value in the
transmitter.
If the operating temperature varies significantly, you must write a temperature value to the
transmitter at appropriate intervals, using an appropriate output assembly. See Section 9.4.
Note: Ensure that your temperature value is accurate, or that your temperature measurement device is
accurate and reliable.
To enable and configure external temperature compensation:
With ProLink II, see Figure 9-3.
With a DeviceNet tool, see Figure 9-4.
Figure 9-3 Configuring external temperature compensation with ProLink II
Configure
Enable Use External
Temperature
Enable
Apply
Enter Temperature units
(1)
Apply
View Menu >
Preferences
ProLink >
Configuration >
Temperature
Set up output
assembly
(2)
Enter External
Temperature
Yes
Done
No
Use static
temp value?
Apply