Satellite Radio User Manual

Configuration and Use Manual 117
Troubleshooting
Measurement Performance DefaultsTroubleshootingCompensation
A009 Transmitter
Initializing/Warming
Up
Transmitter in power-up
mode
Allow the flowmeter to warm up (approximately 30
seconds). The error should disappear once the flowmeter
is ready for normal operation.
If alarm does not clear, make sure that the sensor is
completely full or completely empty.
Check the sensor circuitry. See Section 11.20.
A010 Calibration Failure Mechanical zero: The
resulting zero was greater
than 3 μs.
Temperature/Density Cals:
many possible causes.
If alarm appears during a transmitter zero, ensure that
there is no flow through the sensor, then retry.
Cycle power to the flowmeter, then retry.
If appropriate, restore the factory zero to return the
flowmeter to operation.
A011 Zero Too Low See A10 Ensure that there is no flow through the sensor, then retry.
Cycle power to the flowmeter, then retry.
If appropriate, restore the factory zero to return the
flowmeter to operation.
A012 Zero Too High See A10 Ensure that there is no flow through the sensor, then retry.
Cycle power to the flowmeter, then retry.
If appropriate, restore the factory zero to return the
flowmeter to operation.
A013 Zero Too Noisy See A10 Remove or reduce sources of electromechanical noise,
then retry. Sources of noise include:
- Mechanical pumps
- Pipe stress at sensor
- Electrical interference
- Vibration effects from nearby machinery
Cycle power to the flowmeter, then retry.
If appropriate, restore the factory zero to return the
flowmeter to operation.
A014 Transmitter Failed Many possible causes Cycle power to the flowmeter.
The transmitter might need service. Contact Micro Motion.
See Section 11.3.
A016 Line RTD
Temperature
Out-Of-Range
The value computed for
the resistance of the Line
RTD is outside limits
Check the sensor RTD circuitry. See Section 11.20.
Verify that process temperature is within range of sensor
and transmitter.
If the problem persists, contact Micro Motion. See
Section 11.3.
A017 Meter RTD
Temperature
Out-of-Range
The value computed for
the resistance of the
Meter/Case RTD is
outside limits
Check the sensor RTD circuitry. See Section 11.20.
Verify that process temperature is within range of sensor
and transmitter.
Check the characterization. Specifically, verify the FCF
and K1 values. See Section 6.2.
If the problem persists, contact Micro Motion. See
Section 11.3.
A020 Calibration Factors
Unentered
(FlowCal)
The flow calibration factor
and/or K1 has not been
entered since the last
master reset
Check the characterization. Specifically, verify the FCF
and K1 values. See Section 6.2.
If the problem persists, contact Micro Motion. See
Section 11.3.
A021 Incorrect Sensor
Type (K1)
The sensor is recognized
as a straight tube but the
K1 value indicates a
curved tube, or vice versa
Check the characterization. Specifically, verify the FCF
and K1 values. See Section 6.2.
Check the sensor RTD circuitry. See Section 11.20.
If the problem persists, contact Micro Motion. See
Section 11.3.
A029 PIC/Daughterboard
Communication
Failure
Transmitter electronics
failure
Cycle power to the flowmeter.
Contact Micro Motion. See Section 11.3.
Table 11-2 Status alarms and remedies continued
Alarm
code
ProLink II
message Cause Suggested remedy