ECO-A-X Transmitter Board Terminals
Note: In Table 3, TB1 Terminals (Power Terminals) are non-polarized; if (+) 24VDC is connected
to TB1 bottom terminal, and (-) 24VDC connected to TB1 top terminal a loop current of 4-20mA
would still be obtained.
Sensor separation
Since the sensor must be located where gas is likely to accumulate and the controller where it
can be easily reached, it is often necessary to “separate” the controller and sensor. This is done
with the aid of the Sensor Separation kit. The Sensor Separation kit is composed of a Net Safety
Multi-purpose Junction Box and terminal strip. Refer JB-MPG-A/S in the Multi-purpose
Junction Box manual (MAN-0081) for terminal designations.
TB1 Power Terminals - non polar
TB1
Power
Connections
TB1 top
terminal
(+) 24VDC
TB1 bottom
terminal
(-) 24VDC
TB2 Sensor Terminals
Sensor Wire
Terminal
designation
White
CTR
Red
REF
Black
SEN
Table 2: Sensor Terminal connection
Table 3: Power Terminal connection
TB1 Top
terminal
TB1 bottom
terminal