User's Manual
15
5.5 Sensor Information:
Sensor Information has the SENSOR SETUP/INFO menus shown in Figure 5-7
Install New Sensor should always be performed when a new simple sensor is installed. This
deletes historical CAL data and sets sensor life to 100% after initial calibration of the new simple
sensor.
Recall Cal History recalls each successful calibration. These dates may be reviewed by
scrolling with the UP / DOWN keys.
Input Type indicates what kind of input or sensor the ECO-A-X is configured to accept and is
typically pre-configured at the factory.
New Sensor displays the date when a new sensor was last installed.
Recent Cal displays the most recent calibration date.
5.6 CLOCK/DELAY SETUP:
Since the ECO-A-X is equipped with a Real Time Clock & Calendar Time and Date must be set
to correctly match its location. They are set at the factory in a 24 hour format but may require
adjustment to match the location’s time & date after shipment. Follow the procedure in
Configuration Using the Magnet in section 5.2.