User's Manual

13
Linearization Data allows nonlinear signals to be linearized by entering the correct curve into
the ECO-A-X (Figure 5-4). If Linearize is set for NO, the CURVE POINTS menu data is not used
and no linearization is applied. When YES, the CURVE POINT entries are used and a straight-
line approximation is calculated between each of the 9 entries. 0% input always provides 0%
output and 100% input always provides 100% output. To prevent accidental data entry a special
keystroke sequence, of 4 consecutive UP keys, is required to enter this menu.
Figure 5-4: Linearization Menu
Backup Config. allows users to store the entire current ECO-A-X’s menu database into non-
volatile memory for restoration later if incorrect values are accidentally entered or uploaded.
Restore Config. restores the ECO-A-X’s menu database to the values from the most recent
Backup Config. The special keystroke sequence of 4 consecutive UP keys is also required to
perform backup and restore operations.
TX Sensor Life set for YES, causes the unit’s 4-20mA output to transmit a sensor life value
after successful calibrations during the CAL PURGE delay (see section 4.1). For normal
operation the ECO-A-X transmits 4mA during the CAL PURGE delay. But with the TX Sensor
Life = YES the unit transmits 4mA for the first 10-seconds, then for 5-seconds transmits a value
between 4mA and 5mA, with 4mA equal to 0% sensor life and 5mA equal to 100% sensor life
(see Figure 5-5). The output then returns to 4mA for the remainder of the CAL PURGE delay.
For example, if after a calibration sensor life is 75%, the ECO-A-X transmits 4.75mA during the 5-
second interval.
Note: TX Sensor Life should always be set for NO unless the 4-20mA receiver is capable of
interpreting the sensor life signal.
Figure 5-5: Transmit Sensor Life Timing Diagram
E n t e r K e y S e q u e n c e
t o a c c e s s me n u .
* * * *
NE X T k e y t o E x i t .
S e e Ma n u a l
L i n e a r i z e ? N O
D a t a P o i n t 1
% I n p u t 10 . 0 0
% Ou t p u t 10 . 0 0
PRESS
UP KEY
4-TIMES
C U R V E P OI N T S
Monitoring
Calibrating Cal Delay
Sensor Life Value
4mA=0% life
5mA=100% life
Monitoring
4mA for 10 seconds after Cal Exit Then, 4 - 5mA for 5 seconds
Cal level = 1.5mA if 3-wire transmitter
Cal level = 3.0mA if 2-wire transmitter
4mA