User manual
Emerson Process Management GmbH & Co. OHG 9-11
X-STREAM X2
Instruction Manual
HASX2E-IM-HS
02/2012
9
Modbus Commands
9.3 List of Parameters and Registers - Sorted by Tag Name
Tag Name
Address
Data
Type
Client
Access
Description
Modicon Daniel
Setup.Calibration.RestoreCalibSpan2 2018 2018 Boolean R/W
restore span calibration parameters
from user memory for comp2
Setup.Calibration.RestoreCalibSpan3 2019 2019 Boolean R/W
restore span calibration parameters
from user memory for comp3
Setup.Calibration.RestoreCalibSpan4 2020 2020 Boolean R/W
restore span calibration parameters
from user memory for comp4
Setup.Calibration.RestoreCalibZero1 2013 2013 Boolean R/W
restore zero calibration parameters
from user memory for comp1
Setup.Calibration.RestoreCalibZero2 2014 2014 Boolean R/W
restore zero calibration parameters
from user memory for comp2
Setup.Calibration.RestoreCalibZero3 2015 2015 Boolean R/W
restore zero calibration parameters
from user memory for comp3
Setup.Calibration.RestoreCalibZero4 2016 2016 Boolean R/W
restore zero calibration parameters
from user memory for comp4
Setup.Calibration.SpanGasValue1
6121…
6122
9061 Float R/W
value for comp1 (in ppm) which a span
calibration adjusts to
Setup.Calibration.SpanGasValue2
6123…
6124
9062 Float R/W
value for comp2 (in ppm) which a span
calibration adjusts to
Setup.Calibration.SpanGasValue3
6125…
6126
9063 Float R/W
value for comp3 (in ppm) which a span
calibration adjusts to
Setup.Calibration.SpanGasValue4
6127…
6128
9064 Float R/W
value for comp4 (in ppm) which a span
calibration adjusts to
Setup.Calibration.ToleranceCheck 3021 3021 Word R/W
check deviation tolerance (0=Off,1=On/
AutoClear,2=On)
Setup.Calibration.Valves.SpanValveAssign1 3036 3036 Word R/W assigns span gas of comp1 to valves
Setup.Calibration.Valves.SpanValveAssign2 3037 3037 Word R/W assigns span gas of comp2 to valves
Setup.Calibration.Valves.SpanValveAssign3 3038 3038 Word R/W assigns span gas of comp3 to valves
Setup.Calibration.Valves.SpanValveAssign4 3039 3039 Word R/W assigns span gas of comp4 to valves
Setup.Calibration.Valves.ZeroValveAssign1 3032 3032 Word R/W assigns zero gas of comp1 to valves
Setup.Calibration.Valves.ZeroValveAssign2 3033 3033 Word R/W assigns zero gas of comp2 to valves
Setup.Calibration.Valves.ZeroValveAssign3
3034 3034 Word R/W assigns zero gas of comp3 to valves
Setup.Calibration.Valves.ZeroValveAssign4 3035 3035 Word R/W assigns zero gas of comp4 to valves
Setup.Calibration.ZeroGasValue1
6101…
6102
9051 Float R/W
value for comp1 (in ppm) which a zero
calibration adjusts to
Setup.Calibration.ZeroGasValue2
6103…
6104
9052 Float R/W
value for comp2 (in ppm) which a zero
calibration adjusts to
Setup.Calibration.ZeroGasValue3
6105…
6106
9053 Float R/W
value for comp3 (in ppm) which a zero
calibration adjusts to
Setup.Calibration.ZeroGasValue4
6107…
6108
9054 Float R/W
value for comp4 (in ppm) which a zero
calibration adjusts to
Setup.Communication.SIntModbusFt32 3026 3026 Word R/W
32bit regs format (0=Daniel, 1=Modi-
con, 2=Modicon swap)