84 Using a Field Communicator to configure the meter
Section 4: Configuration Daniel 3814 Liquid Ultrasonic Meter Installation Manual
March 2015 3-9000-760 Rev E
5. Wire Analog Input 1 (AI1) and Analog Output 1 (AO1) as shown in Figure 4-20 and
Appendix A, drawing DMC-005324.
Figure 4-20 Field Communicator wiring diagram for the 3810 Series electronics
6. Use the leads provided with the Field Communicator to connect to your device.
7. Press and hold the
Power button on the Field Communicator until the green light blinks.
8. Use the touch screen on the Field Communicator, the keypad or use the stylus to
navigate through the device menus.
9. Refer to the Menu tree in Section D.1.1 of the Daniel HART Field Device Specification
manual (P/N 3-9000-762) for the device fast key sequences. Included in the menu tree
are:
Diagram Page 1 - 3410 Series Root Menu; Overview, Configure>Manual Setup
Diagram Page 2 - Configure>Manual Setup (continued) and Alerts Setup
Diagram Page 3 - Service Tools>Alerts and Variables
Diagram Page 4 - Service Tools>Variables (continued), Service Tools>Trends, and
Service Tools>Maintenance
If you encounter problems, refer to the contact information on the back cover of this manual or
the contacts included in the Field Communicator User’s Manual.