User Manual

Daniel 3814 Liquid Ultrasonic Meter Installation Manual Section 4: Configuration
3-9000-760 Rev E March 2015
Field Setup Wizard 67
Display units
The Meter volume units displayed are either U.S. Customary or Metric. To modify the Display
Units, configure the Meter units system in the Field Setup Wizard — General Page.
U.S. Customary volume unit selections are:
Barrels
Gallons
Metric volume unit selections are:
Cubic meters
Cubic liters
Display units preceded by a plus or minus sign indicate forward and reverse flow
direction, as shown in the table below.
The Local Display Flow rate time units are modifiable by selecting the drop-down
arrow and clicking the time unit in the list box.
Valid flow rate time units selections are:
second
minute
hour
day
Scroll delay
The Scroll Delay is the time interval for the selected display items to be shown on the Local
Display. The default scroll delay setting is five seconds. Click the spin box up or down arrow to
increase or decrease the length of time an item displays.
1. Select
Finish to write the configuration settings to the meter.
2. Save the meter configuration file, collect a Maintenance log and Waveforms to
document the “As Left” settings.