Daniel 3814 Liquid Ultrasonic Meter Installation Manual Section 2: Mechanical installation
3-9000-760 Rev E March 2015
Piping recommendations 27
Refer to the ultrasonic meter product data sheet for piping information. The Liquid Ultrasonic
Flow Meter Datasheet may be downloaded from the Daniel website:
http://www2.emersonprocess.com/en-US/brands/daniel/Flow/ultrasonics/Pages/Ultrasonic-Series-
3800.aspx
Meter tube dimensions with tube bundle or profiler plate for uni-directional and bi-directional
flow, the minimum straight pipe length is as follows:
The bore of the mating piping should be within 1% of the meter inside diameter.
The meter is provided with dowel pins to align the meter body bore with the bore of the
mating piping.
The Daniel Liquid Ultrasonic Flow Meter should be mounted in horizontal piping with the
chord paths horizontal
Normally, the meter body is installed so that the electronics assembly is on the top of the
meter. If there is insufficient space above the piping for this arrangement, the meter can
be ordered with extra long transducer cables for remote mounting or the meter housing
can be installed with the electronics assembly on the bottom.
The mating piping should include temperature and pressure measurement connections
located a minimum of two nominal pipe diameters length down stream of the meter, or
per API MPMS 5.8.
Piping requirements for field calibration of high viscosity, low Reynolds Number
applications is comprised of the upstream pipe section, CPA (Canadian Pipeline
Associates) flow conditioning plate, Venturi reducer section, ultrasonic meter and
downstream pipe section.
Table 2-1 Piping recommendation for uni-directional or bi-directional flow
Uni-Directional Flow Bi-Directional Flow
8D up stream
(with a flow conditioner)
8D up stream
(with a flow conditioner)
8D up stream
(no flow conditioner)
8D up stream
(no flow conditioner)
5D down stream 8D down stream
To access the product datasheet, from the Daniel products page (above link), select the
Daniel Liquid Ultrasonic Flow Meter link, click the Documentation tab, expand the Data
Sheets - Bulletins - Catalogs tab, then select the Data Sheet.
FAULTY METER INSTALLATION
Correctly install the equipment
If meter bodies are mounted or oriented differently than specified above, debris or gas may
collect in the transducer ports which could adversely affect the transducer signals, or cause
equipment damage.