12 Meter specifications
Section 1: Introduction Daniel 3814 Liquid Ultrasonic Meter Installation Manual
March 2015 3-9000-760 Rev E
Electronic specifications
Power Meter
• 10.4 VDC to 36 VDC
• 11 W typical power consumption
Serial cable
• Belden #9940 or equivalent (22 gauge)
– Capacitance (pF/m) 121.397 (conductor to conductor)
– Capacitance (pF/m) 219.827 (conductor to other conductor and shield)
– Resistance (DC) DCR @ 20 °C (Ohm/km) 48.2307
Nominal Outer shield resistance - DCR @ 20 °C (Ohm/km) 16.405
– Operating voltage - 300 V RMS (UL AWM Style 2464)
– Current 2.4 Amps per conductor @ 25 °C (recommended)
Ethernet cable
• Cat-5 Standard 100Mbps
Frequency (
see Table 3-1)
• 22 AWG wire characteristics are as follows:
– Capacitance = 20 pF/ft or 20 nF/1000 ft (between two wires)
– Resistance = 0.0168 Ohms/ft or 16.8 Ohms/1000 ft
– Pull-up voltage is 24 VDC
Temperature
flameproof Transmitter
Electronic Enclosure and
Base Electronics Enclosure
Transducers
• LT-01, LT-03
– Operating Temperature Range -58
o
F to +212
o
F (-50
o
C to 100
o
C)
• LT-04, LT-05, LT-08, LT-09
– Operating Temperature Range -58
o
F to +302
o
F (-50
o
C to +150
o
C)
Note: The process temperature must not exceed the operating temperature range of
the transducers.
Note: The ultrasonic transducers are not intended for use across boundary walls of different
hazardous area classifications. The transmitter electronics cannot be remote mounted
from a Division 1 classification to a Division 2 area to meet an area classification.
Communications specifications
Connectivity protocols One serial RS-232/RS-485 ports (115 kbps baud rate) (Modbus RTU/ASCII)
• (1) Serial Port A
(RS-232/RS-485 Full Duplex/RS-485 Half Duplex)
One Ethernet Port (TCP/IP) 100 BaseT
• Up to 10 Mbps
• Modbus TCP
Device compatibility FloBoss 103, FloBoss S600 flow computer, ROC 107,
Digital, analog, and frequency inputs
Digital Input(s)
(Selectable)
(1) Single polarity (for flow calibration gating - contact closure)
– Single input for starting and stopping
– Four pulse configurations available
Analog Input(s) (2) 4-20 mA
• AI-1 Temperature
• AI-2 Pressure
Note: The analog-to-digital conversion accuracy is within ±0.05% of full scale over the operating
temperature range.
Note: AI-1 and AI-2 are electronically isolated and operate in sink mode. The input contains a
series resistance so HART® Communicators can be connected to configure sensors.
A 24 Volt DC power supply is available to provide power to the sensors.
Table 1-2 Meter specifications