User's Manual
Table Of Contents
- 1 Safety Directions
- 2 Description of the System
- 3 Accessories
- 4 Operation
- 5 Check & Adjust
- 6 Care and Transport
- 7 Technical Data
- 8 Software Licence Agreement/Warranty
- Appendix A Abbreviations
- Appendix B Regional Contact Addresses
•
Angle measurements
•
Distance measurements
•
Reectors
2.8.1 Maximum Permissible Error (MPE)
Accuracy specications in the technical data of this manual are stated with
the Maximum Permissible Error (MPE). The Maximum Permissible Error (MPE)
is dened by the ISO 10360-10:2021 as the extreme value of the measure-
ment error permitted by specication for the relevant test, i.e. length meas-
urement. An acceptance test is veried if all measurement errors according
to the relevant test item are smaller than the corresponding MPE specic-
ations. The ISO 10360-10:2021 denes criteria for the repetition of meas-
urements that exceed the MPE values. Typical measurement results of the
Absolute Tracker are half of the relevant MPE values.
2.8.2 Laser Tracker Subsystem Specications
The following specications represent the Laser Tracker subsystem specica-
tion in accordance with ISO 10360-10:2021, Annex D.
Laser Tracker Subsystem Symbol Maximum Permissible Error
(MPE)
Absolute Distance Meter
(ADM)
e
ADM
±10 μm + 0.3 μm/m
R0 Parameter (R0) e
R0
±5 μm
Transverse e
T
±15 μm + 6 μm/m
2.8.3 Measurement Uncertainty
The table below shows the Maximum Permissible Error (MPE) specications
for the test items dened in ISO 10360-10:2021 using Leica 1.5'' Red Ring
Reectors (RRR) and standard measurement mode unless otherwise noted.
Two-Face Meas-
urement
Position Tolerance
(MPE)
General Formula
Absolute Angular
Performance e
T
1)
Pos. 1 to 9
(1.5 m dis-
tance)
±24 μm ±15 μm + 6 μm/m
Pos. 10 to
18
(6 m dis-
tance)
±51 μm
1)
Angular Performance Transverse e
T
according to ISO 10360-10:2021,
this respects to a MPE for the Location Error (LDia.2×1:P&R:LT,MPE) in
accordance with chapter 6.3 of ISO 10360-10:2021 of 30 μm + 12 μm/m
Description
Description
Description of the System 39