Currents
As mentioned previously, after you have Plugged in your Current Probes it will be
automatically recognized. Therefore, only minor configuration adjustments will
need to be made in the Current Configuration Window.
Access your Portable BLACKBOX Unit via Elspec's Web Interface log on
as the Administrator under Configuration Device Setup select the
Currents Tab:
In the Currents Window adjust:
Primary & Secondary Transformation Ratios for all the Current
channels,
Nominal Ampere Values for all the Current Channels (Nominal is set for
the current measurements that will define PQZIP threshold / tolerance
value & it is also used for event settings)
Either Reverse the Polarity / maintain it at Normal from the drop-down
selection (Polarity toggling is used to correct incorrect wiring)
Clamp’s Current / Voltage Ratio (As per Clamp's individual
Specifications) &
Hardware Range (As per Clamp's individual Specifications):