User guide
M-Series Touchcomputer User Guide xiii
Recommended separation distances between portable
and mobile RF
communications equipment and the 15MX/19MX Touchcomputer for Healthcare
Applications for all EQUIPTMENT AND SYSTEMS that are not
LIFE-SUPPORTING
The 15MX/19MX Touchcomputer for Healthcare Applications is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The customer or the user of the
15MX/19MX Touchcomputer for Healthcare Applications can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications (equipment)
and the 15MX/19MX Touchcomputer for Healthcare Applications as recommended below according to
the maximum output power of the communications equipment.
Separation Distance According to Frequency of Transmitter
Rated Maximum
Output Power of
Transmitter
(W)
150 kHz to 80 MHz
80MHz to 800 MHz
800 MHz to 2.5 GHz
0.0
1
0.1
2
0.1
2
0.23
0.
1
0.3
7
0.3
7
0.74
1 1.
2
1.
2
2.3
13. 3. 7.4
10
0
1
2
1
2
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distanced in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.