Guidance and manufacturer’s declaration-electromagnetic
immunity
for all EQUIPMENT AND SYSTEMS
Guidance and manufacturer’s declaration-electromagnetic immunity
The 15MX/19MX Touchcomputer for Healthcare Applications is intended for use in the electromagnetic
environment specified below. The customer, or the user of the 15MX/19MX Touchcomputer for Healthcare
Applications, should assure that it is used in such an environment.
Immunity Test IEC 60601 Compliance Level
Electromagnetic Environment-
Guidelines
Electrostatic Discharge
(ESD)
IEC 61000-4-2
±
6 kV contact
±
8 kV air
±
6 kV contact
±
8 kV air
Floors should be wood,
concrete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
Electrical Fast
Transient/Burst
IEC 61000-4-4
±
2 kV for power
supply lines
±
1 kV for input/outpu
t
lines
2 kV for power supply
lines
±
1 kV for input/outpu
t
lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±
1 kV line(s) to
line(s)
±
2 kV line(s) to earth
±
1 kV line(s) to
line(s)
±
2 kV line(s) to earth
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage Dips, Short
Interruption and Voltage
Variations on Power
Supply Input Lines
IEC 61000-4-11
<5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 250 cycles
<5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60 % dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 250 cycles
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the 15MX/19MX Touchcompute
r
for Healthcare Applications
requires continued operation
during power mains interruptions,
It is recommended that the
15MX/19MX Touchcomputer for
Healthcare Applications be
powered from an uninterruptible
power supply or a battery.
Power Frequency
(50/60 Hz) Magnetic
Field
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
NOTE UT is the A.C. mains voltage prior to application of the test level.