C-Series Touchcomputer for Healthcare Applications xii
3. Field strengths from
fixed
transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to
fixed
RF transmitters, an electromagnetic site survey should be
considered. If the measured
field
strength in the location in which the 19CX/22CX
Touchcomputer for Healthcare Applications is used exceeds the applicable RF compliance
level above, the 19CX/22CX Touchcomputer for Healthcare Applications should be observed
to verify normal operation. If abnormal performance is observed, additional measures may
be necessary, such as reorienting or relocating the 19CX/22CX Touchcomputer for
Healthcare Applications.