User manual
© 2014 Elo Touch Solutions, Inc. All rights reserved.
User Manual – 1929LM
SW200158 Rev A - Page 8 of 31
Guidance and manufacturer’s declaration-electromagnetic immunity –
for all ME EQUIPMENT AND ME SYSTEMS
Guidance and manufacturer’s declaration-electromagnetic immunityError! Bookmark not
defined.
The 1929LM Touchmonitor for Healthcare Applications is intended for use in the electromagnetic
environment specified below. The customer, or the user of the 1929LM Touchmonitor for Healthcare
Applications, should assure that it is used in such an environment.
Immunity Test
IEC 60601
test level
Compliance level
Electromagnetic environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 2 kV for power
supply lines
± 1 kV for
input/output Lines
Mains power quality should be that of a
typical commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Mains power quality should be that of a
typical commercial or hospital environment.
Voltage dips, short
interruption and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 250 cycles
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60 % dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 250 cycles
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the 1929LM Touchmonitor for
Healthcare Applications requires continued
operation during power mains interruptions,
It is recommended that the 1929LM
Touchmonitor for Healthcare Applications be
powered from an uninterruptible power
supply or a battery.
Power Frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location in
a typical commercial or hospital
environment.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.