User manual
© 2014 Elo Touch Solutions, Inc. All rights reserved.
User Manual – 1929LM
SW200158 Rev A - Page 5 of 31
Classification
With respect to electrical shock, fire in accordance with ANSI/AAMI ES60601-1:2005 and
CAN/CSA C22.2 No. 60601-1-08
This Touchmonitor is a Class I (GROUNDED) DEVICE.
These Touchmonitors are classified NO APPLIED PARTS EQUIPMENT.
Protection against harmful ingress of water:
INGRESS PROTECTION (IPX1)
This Touchmonitor shall be classified as ORDINARY EQUIPMENT, not intended or evaluated
for use in the presence of flammable anesthetic mixture with air, oxygen, or nitrous oxide.
Mode of Operation: CONTINUOUS OPERATION.
Environmental conditions for transport and storage
Temp.
Operating
0°C to 40°C
Storage / Transportation
-20°C to 60°C
Humidity (non-condensing)
Operating
20% to 80%
Storage / Transportation
10% to 90%
Altitude
Operating
0 to 3,048m
Storage / Transportation
0 to 12,192m
1929LM Touchmonitor for Healthcare Applications is intended for general use in hospital environment for data
collection and display for reference. It shall not be used with life-supporting system.