User Manual

Trouble shooting
© Elma GmbH & Co KG BA/Elmasonic S/GB/0207
31
Fault Possible cause Remedy
turning knob temperature in
„0“ position
switch on turning knob
temperature
unit is switched off switch on unit with key on/off
Unit does not heat up;
LED temperature dark
fault of electronics return unit to supplier or
manufacturer
No heating function;
LEDs of LED temperature
blink alternately ("running
light") = fault indication
heating
fault of electronics switch unit off and on
if fault is indicated again:
return unit to supplier or
manufacturer
loss of heating energy use cover (optional
accessory equipment)
Unsatisfactory heating-up
period
no mixing of cleaning liquid e.g. switch on ultrasound
(see section 7.2)
Unit produces boiling
noise during heating-up
no mixing of cleaning liquid e.g. switch on ultrasound
(see section 7.2)
temperature sensor does not
measure the average
temperature (no revolution)
mix liquid manually or by
means of ultrasound
for low set temperatures do
not switch on heating
Set temperature is
exceeded
set temperature too low,
ultrasonic energy heats up the
liquid more than required
(physical process)
switch on ultrasound for
short periods only
No operational functions;
LEDs of LED ultrasound
and LED temperature blink
alternately ("running light")
= fault indication
programme control
fault of electronics switch unit off and on
if fault is again indicated:
return unit to supplier or
manufacturer