11 Trouble shooting
Fault Possible cause Remedy
Casing damaged damage by third party,
transport damage
return unit to supplier or
manufacturer
Mains cable damaged damage by third party,
transport damage
obtain original spare mains
cable from manufacturer or
supplier
mains cable not plugged in plug in mains cable
socket dead check socket/fuse
mains cable
damaged/interrupted
replace mains cable
No operating functions; all
LEDs dark
fault of electronics return unit to supplier or
manufacturer
turning knob for ultrasonic
operation in „0“ position
switch on the turning knob
for ultrasonic operation
unit is switched off switch on the unit at key
on/off
key (ultrasound) not
pressed
press key
No ultrasonic function;
LED ultrasound dark
fault of electronics return unit to supplier or
manufacturer
adverse filling level change filling level, switch
unit and off and on
No ultrasonic operation;
LEDs of LED cleaning
period blink alternately
("running light") = fault
indication ultrasound
fault of electronics switch unit off and on
if fault is indicated again:
return unit to supplier or
manufacturer
no or unsuitable cleaning
medium used
use suitable cleaning
medium
cleaning temperature not
sufficient
heat up cleaning liquid
Unsatisfactory cleaning
results
cleaning period too short
repeat cleaning interval