User's Manual

Chapter 3 EMC TEST REPORT
3.1 Description of Standards and Results for EN55022/EN55024
EMISSION
Description of Test Item
Standard
Limits
Results
Conducted disturbance
CISPR 11:2009+A1:2010
EN 55011:2009+A1:2010
Group 2, Class B
PASS
Minimum passing margin is 6.69dB at
0.158MHz
Radiated disturbance
(30-1000MHz)
CISPR 11:2009+A1:2010
EN 55011:2009+A1:2010
Group 2, Class B
PASS
Minimum passing margin is 3.10dB at
593.628MHz
Radiated disturbance
(Above 1GHz)
CISPR 11:2009+A1:2010
EN 55011:2009+
A1:2010
Group 2, Class B
PASS
Minimum passing margin is 3.22dB at
1482.48MHz
Harmonic current emissions
IEC 61000-3-2:2014
Class D
PASS
Voltage fluctuations & flicker
IEC 61000-3-3:2013
Section 5
PASS
IMMUNITY (EN 60601-1-2 2015)
Description of Test Item
Standard
Limits
Results
Electrostatic discharge (ESD)
IEC 61000-4-2: 2008
Section 6.2.1.10
PASS
Radiated RF electromagnetic
fields
IEC 61000-4-3: 2010
Section 6.2.1.10
PASS
Electrical fast transients and
bursts
IEC 61000-4-4: 2012
Section 6.2.1.10
PASS
Surge
IEC 61000-4-5: 2005
Section 6.2.1.10
PASS
Conducted disturbances,
induced by RF fields
IEC 61000-4-6: 2013
Section 6.2.1.10
PASS
Power frequency magnetic fields
IEC 61000-4-8: 2009
Section 6.2.1.10
PASS
Voltage dips
IEC 61000-4-11:2004
Section 6.2.1.10
PASS
Voltage interruptions
Section 6.2.1.10
PASS
NOTE
N/A is an abbreviation for Not Applicable
NOTE: means a deviation from the requirement of EN 60601-1-2 Section 6.2.1.10), but it’s allowed.
mPAD2-7-CHT4-I Tablet PC User Manual 17