Specifications

ELGA PURELAB Option-R 7/15 Operator Manual
PURELAB Option-R 7/15 Version 4 08/07 Page 21
6.2 Alarm Conditions
Alarms will signal at the following conditions:
Replace Ion-exchange Cartridge
The ion-exchange cartridge replacement alarm is signalled by an
audible alarm and flashing icon at the default setting of 4380 hrs
(6 months) of use. Press the
button to mute the audible alarm.
Follow the instructions to replace the ion-exchange cartridge. (See
Section 7.2).
Replace UV Lamp
The UV lamp replacement alarm is signalled by an audible alarm and
flashing icon at the default setting of 8760 hrs (1 year) of use. Press
the
button to mute the audible alarm. Follow the instructions to
replace the UV lamp. (See Section 7.3).
Replace Pre-treatment Cartridge
The pre-treatment cartridge replacement alarm is signalled by an
audible alarm and flashing icon at the default setting of 4380 hrs (6
months) of use. Press the
button to mute the audible alarm. Follow
the instructions to replace the pre-treatment cartridge. (See Section
7.1).
UV failure
The UV failure alarm is signalled by an audible alarm and flashing
cross over the replace UV icon. Press the
button to mute the
audible alarm. Follow the instructions to replace the UV lamp. (See
Section 7.3).
The PURELAB Option-R will continue to run without the UV lamp
operating.
CAUTION! Long term operation without the UV lamp
in operation will compromise performance
and is not recommended.
Low Level Alarm
When the low level alarm sounds, the mimic reservoir on the display
will flash and a crossed bell icon mute symbol will appear. To mute
the low level alarm sound PRESS the
button. The PURELAB
Option-R will automatically refill the reservoir.
Water Purity Alarm
This alarm will signal if the water purity deviates from the preset
parameters and will cause the water purity to flash and an alarm to
sound, until water purity improves to within acceptable purity limits.
Press the
button to mute the alarm. If water purity stays outside
acceptable limits replace the ion-exchange pack following the
instructions in section 7.2.
Flash
Mute Icon
Low Level Alarms / muted
Flash
Flash
Water Purity Alarm / Standby
Alarm Conditions