Specifications

ELGA PURELAB Option-R 7/15 Operator Manual
PURELAB Option-R 7/15 Version 4 08/07 Page 13
Outlet to
reservoir
Recirculation
from reservoir
Blank
Drain
Recirculation and Drain Connections
5.3 Connecting up the PURELAB Option-R
Once the PURELAB Option-R unit has been positioned either on a
wall or on a bench, it should be connected as follows:
Mains water inlet tube
Drain
Re-circulation from reservoir
Outlet to reservoir
Step 1 - Fitting Tubes
1. PUSH in collet on connector.
2. PULL out transit plug.
3. CUT a clean square end on a 8mm (5/16") OD
semi rigid drain tube.
4. PUSH tube into connector.
CAUTION! Do not restrict drain line.
CAUTION! If the water supply is at a pressure greater
than 6 bar (90 psi) fit a pressure regulator
(LA512).
Connecto
r
Collet
Blanking
transit plug
Pull plug
Depress
collet
Clean square
cut end
Tube 8mm OD
Simply push in tube to attach
Tube in secured position
Fitting Tubes
Feedwater Inlet Connections