User Manual
Table 8. ACV Detailed Accuracy Specifications
Table 9. DCA Detailed Accuracy Specifications
Table 10. ACA Detailed Accuracy Specifications
Table 11. Resistance Detailed Accuracy Specifications
-34-
Range Resolution Accuracy Note
400mV 0.1mV ±(1.5% rdg + 8) Frequency Response:
50Hz - 400Hz
Input Resistance: 10MΩ
Overload Protect: 400mV
range 250V rms. The rest
600VDC or 600VAC rms.
4V 1mV
±(0.8% rdg + 5)
40V 10mV
400V 100mV
600V 1V ±(1.2% rdg + 5)
Range Resolution Accuracy Note
40µA 0.1µA
±(2.0% rdg + 5)
Overload Protect: Fast fuse
0.5A/250V and fast fuse
10A/250V.
10A for 15sec. maximum
Input Voltage Drop: <0.4V
400µA 1µA
40mA 10µA
±(1.5% rdg + 5)
400mA 100µA
4A 1mA
±(2.0% rdg + 5)
10A 10mA
Range Resolution Accuracy Note
40µA 0.1µA
±(2.5% rdg + 3)
Overload Protect: Fast fuse
0.5A/250V and fast fuse
10A/250V.
10A for 15sec. maximum
Input Voltage Drop: <
0.4V
Frequency Response:
50Hz - 400Hz
400µA 1µA
40mA 10µA
±(2.0% rdg + 5)
400mA 100µA
4A 1mA
±(2.5% rdg + 5)
10A 10mA
Range Resolution Accuracy Note
400Ω 100mΩ
±(1.0% rdg + 5)
Overload Protect: 250V rms
4kΩ 1Ω
40kΩ 10Ω
400kΩ 100Ω
4MΩ 1kΩ
40MΩ 10kΩ ±(2.0% rdg + 5)