PSB 9000 2.5 kW Series
IF Level 1 Description
RS232
-
Thebaudrateisselectable,otherserialsettingscan’tbechangedandaredenedlikethis:8
data bits, 1 stop bit, parity = none
Baud rates: 2400, 4800, 9600, 19200, 38400, 57600, 115200
Element Description
Com Timeout Timeout USB/RS232 (in milliseconds)
Default value: 5, Range: 5...65535
Denesthemax.timebetweentwosubsequentbytesorblocksofatransferredmessage.
For more information about the timeout refer to the external programming documentation
“Programming ModBus & SCPI”.
Timeout ETH (in seconds)
Default value: 5, Range: 5...65535
Denesatimeoutafterwhichthedevicewouldclosethesocketconnectioniftherewasno
command communication between the controlling unit (PC, PLC etc.) and the device for the
adjustedtime.Thetimeoutisine󰀨ectiveaslongasoption“TCPkeep-alive”isenabledand
the keep-alive network service is running.
Com Protocols Enables or disables SCPI or ModBus communication protocols for the device. The change is
immediatelye󰀨ectiveaftersubmittingitwithENTERbutton.Onlyoneofbothcanbedisabled.
Logging Default setting: disabled
Enables/disablesthe“logtoUSBstick”feature.Onceenabled,youcandenetheloginterval
(multiple steps, 500 ms ... 5 s) and the control method. For more refer to „3.5.8. Recording to
USB stick (logging)“.
3.5.3.8 Menu “HMI Setup”
These settings refer exclusively to the control panel (HMI).
Element Description
Language Selection of the display language between German, English (default), Russian or Chinese
Key Sound Activates or deactivates sounds when touching a touch area in the display. It can usefully
signal that the action has been accepted.
Alarm Sound Activatesordeactivatestheadditionalacousticsignalofanalarmoruserdenedeventwhich
has been set to “Action = ALARM”. See also „3.7 Alarms and monitoring“ on page 57.
HMI Lock See „3.8 Control panel (HMI) lock“ on page 59.
Backlight The choice here is whether the backlight remains permanently on or if it should be switched
o󰀨whennoinputviascreenorrotaryknobismadefor60s.Assoonasinputismade,the
backlight returns automatically. Furthermore, the backlight intensity can be adjusted here.
Limits Lock See „3.9 Limits lock“ on page 60