2.3.4.3 Strain relief and plug xture
ThereisastandardxturemountedtotheACinputcon-
nection block on the rear. It’s used to prevent the AC plug
from loosening and disconnecting due to vibrations or
similar.Thextureisalsousedasstrainrelief.
Using the 4x M3 acorn nuts, it’s recommend to mount the
xturetotheAClterblockeverytimetheACplughas
been plugged again.
It’s furthermore recommended to install the strain relief
by using suitable cable straps (1 or 2 pcs, not included),
asdepictedintheguretotheright.
2.3.4.4 Connection variants
In a situation when multiple units are operated in parallel and there is a three-phase supply available, it should be
used in order to distribute the phase load as equally as possible. The ideal setup would be three units or a multiple
ofthree.Anynumberofunitsonthesameconnectionpointresultsinadierentconnectionscheme.Herearea
few example for comparison:
Single unit Two units, unbalanced phase load
PSB
2.5 kW
L1(A)
N
L2
L3
L1
(A)
L2
(B)
L3
(C)
N
PSB
2.5 kW
L2
L3
PSB
2.5 kW
L2
L3
oder
PSB
2.5 kW
L2(B )
N
L2
L3
oder
PSB
2.5 kW
L3(C )
N
L2
L3
Three units, balanced phase load
L1
(A)
L2
(B)
L3
(C)
N
PSB
2.5 kW
L2
L3
PSB
2.5 kW
L2
L3
PSB
2.5 kW
L2
L3