PSB 9000 2.5 kW Series
3.5.7 Switching the DC terminal on or o
TheDCterminalofthedevicecanbemanuallyorremotelyswitchedonando.Thiscanberestrictedinmanual
operation by the control panel being locked. After switching it on, it either works as input (sink mode) or output
(source mode). More information can be found in „3.3.6. Sink-source mode switching“.
Switching the DC terminal on during manual operation or digital remote control can be disabled
by pin REM-SB of the built-in analog interface. For more information refer to 3.5.3.1 and ex-
ample a) in 3.6.4.7.
► How to manually switch the DC terminal on or o
1. As long as the control panel isn’t fully locked press the button On/O. Otherwise you are asked to disable
theHMIlockrst.
2. ThisbuttontogglesbetweenDCterminalonoro,aslongasachangeisn’trestrictedbyanalarmorthe
device being locked in “remote”.
► How to remotely switch the DC terminal on or o via the analog interface
1. See section “„3.6.4 Remote control via the analog interface (AI)“ on page 52.
► How to remotely switch the DC terminal on or o via the digital interface
1. See the external documentation “Programming Guide ModBus & SCPI” if you are using custom software,
or refer to the external documentation from LabView VIs or other software provided by the manufacturer.
3.5.8 Recording to USB stick (logging)
Device data can be recorded to USB stick (2.0 / 3.0 may work, but not all vendors are supported) anytime. For
specicationsoftheUSBstickandthegeneratedloglesrefertosection„1.9.6.5. USB port (front side)“.
TheloggingstoreslesofCSVformatonthestick.ThelayoutofthelogdataisthesameaswhenloggingviaPC
with software EA Power Control. The advantage of USB logging over PC logging is the mobility and that no PC is
required.TheloggingfeaturejusthastobeactivatedandconguredintheMENU.
3.5.8.1 Conguration 1
Also see section 3.5.3.7. After USB logging has been enabled and the parameters “Logging interval” and “Start/
Stop” have been set, logging can be started anytime from within the MENU or after leaving it, depends on the
selected start/stop mode.
3.5.8.2 Conguration 2
Also see section 3.5.3.1.ThereareadditionalsettingsfortheCSVleitselfasgeneratedbytheUSBloggingfea-
tures. You can switch the column separator format between german/european standard (“Default”) or US american
standard (“US”). The other option is used to deactivate the physical unit that is added by default to every values
inthelogle.DeactivatingthisoptionsimpliestheCSVleprocessinginMSExcel.
3.5.8.3 Handling (start/stop)
With setting “Start/stop with DC terminal ON/OFF” logging will start each time the DC terminal of the device is
switchedon,nomatterifmanuallywiththefrontbutton“On/O”orremotelyviaanalogordigitalinterface.With
setting “Manual start/stop”it’sdierent.LoggingisthenstartedandstoppedonlyintheMENU,inthelogging
congurationpage.
Soon after logging has been started, the symbol indicates the ongoing logging action. In case there is an error
while logging, such as USB stick full or removed, it will be indicated by another symbol ( ). After every manual
stoporswitchingtheDCterminalotheloggingisstoppedandthelogleclosed.