PS 9000 2U Series
3.3.7 The quick menu
The quick menu is an alternative menu for quick access to not directly adjustable set values such as power (in
display mode UI) or current (in display mode UP(I)). It is only available while the DC output is switched on.
The quick menu is accessible with the button and looks like this:
Navigation in the menu is also done with arrow buttons / and .
The power set value P can be accessed, for example, with three short button actions, such as the HMI lock. It is
important to point out, that the set values in submenu “Output Settings” have to be submitted with button .
3.3.8 Switching the DC output on or off
The DC output of the device can be manually or remotely switched on and off. This can be restricted in manual
operation by the control panel being locked.
► How to manually switch the DC output on or off
1. As long as the control panel (HMI) is not fully locked press the button ON/OFF. Otherwise you are asked to
disabletheHMIlockrst.
2. This button toggles between on and off, as long as a change is not restricted by an alarm or the device is
locked in “remote”. The current condition is displayed as “Output ON” or “Output OFF”.
► How to remotely switch the DC output on or off via the analog interface
1. See section “„3.4.4 Remote control via the analog interface (AI)“ on page 49.
► How to remotely switch the DC output on or off via the digital interface
1. See the external documentation “Programming Guide ModBus & SCPI” if you are using custom software,
or refer to the external documentation of LabView VIs or other software provided by the manufacturer.