User Manual
12
Instruction manual
PS 2000 B Single Series
EN
Date: 06-03-2011
Operating the device
Pushbutton „On/Off“
Is used to switch the DC output on or off.
Power output, safety sockets, poled
The sockets can be used to plug 4mm open or safety
Bueschel plugs. The metal socket between the DC
output sockets is a grounding contact and can be used
to ground connected loads, if required.
5. Handling
5.1 The display
Figure 3
5.1.1 Status tokens
The status tokens in the display indicate following:
CV - Voltage regulation active (only if output is „on“)
CC - Current regulation active (only if output is „on“)
Preset V/C - Set value display of voltage/current active
Preset OVP/OCP - Set value display of OVP/OCP active
OT - Overtemperature error
OCP - Overcurrent protection
OVP - Overvoltage protection
Remote - Remote control active (via USB)
Lock - Control panel lock active
5.1.2 Error indication
If an error like overvoltage, overcurrent or overtempera-
ture occurs it is displayed by the text „Error“ and a token
(OT, OCP, OVP) and the output voltage is cut off. The text
remains in the display until the user has acknowledged
the error with the „On/Off“ button, which will also switch
the output off. After an overtemperature error, the output
voltage will return automatically and „Error“ will be cleared,
unless the output has been switched off by the user. Other
errors require the user to switch the output on again, in
order to continue working with the device.
Other display elements are connected to certain operation
modes and are explained in the following sections.
5.2 Pushbuttons
5.2.1 Pushbutton Preset
This button is used to switch to preset mode and for acti-
vation/deactivation of the LOCK mode.
Push Display Mode
1x
Display of U/I set values
2x
Display of OVP/OCP set
values
3x
Activation/Deactivation of
LOCK mode (also see 5.5)
4x Display of U/I actual values
Set values are not stored when switching off the de-
vice and after the next start they are reset to 0% voltage
and 100% current.
5.2.2 Pushbutton On/Off
This pushbutton is used to manually switch the power out-
put on or off, as long as the device is not in remote control.
The state of the output is indicated by
or in the display.
The pushbutton may be locked by the LOCK state. See
above. Also see „5.5 Control panel lock (LOCK)“.
The button also acknowledges errors. See section 5.1.2
for details.
5.3 Further control elements
Knobs Voltage & Current
These two knobs have no stop and are used to adjust set
values. Assignment:
• Left knob - Voltage (U) or OVP, depending on mode
• Right knob - Current (I) or OCP, depending on mode
For details read below.