User Manual
Page 45
EA Elektro-Automatik GmbH
Helmholtzstr. 31-33 • 41747 Viersen
Germany
Fon: +49 2162 / 3785-0
Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
ELR 9000 Series
3.3.3.8 Menu “HMI settings”
These settings refer exclusively to the control panel (HMI).
Element Description
Language Selection of the display language. Currently (date: 06-13-2013) available are:
German, English
Additional languages (up to three can be integrated) can be implemented on demand
and installed into the HMI via update. The default then would be English plus one or two
additional language such as Italian, Spanish etc.
Backlight Setup The choice here is whether the backlight remains permanently on or if it should be
switched off when no input via screen or rotary knob is made for 60s. As soon as input
is made, the backlight returns automatically. Furthermore the brightness can be selected
in 10 steps.
Panel Contrast Here the user can adjust the display contrast to suit best.
HMI Lock See „3.6 Control panel (HMI) lock“ on page 54.
Key Sound Activates or deactivates sounds when touching a touch area in the display. It can usefully
signal that the action has been accepted.
Alarm Sound Activates or deactivates the additional acoustic signal of an alarm or user dened event
which has been set to “Action = ALARM”. See also „3.5 Alarms and monitoring“ on page 52.
HMI Update With this function the rmware of the control panel can be updated using a USB ash
drive. For details refer to „4.3.1 HMI update“ on page 73.
TP Calibration Starts a calibration process for the touchscreen. After calibration the touch areas should
react better and more accurately to nger touches. Internal settings are saved, but a
general reset of the device will reset them to the factory defaults and a recalibration will
be needed.
3.3.4 Adjustment limits
Adjustment limits are only effective in manual operation and only on the set values in the
main display screen!
Defaults are that all set values (U, I, P, R) are adjustable from 0 to
100%..
This may be obstructive in some cases, especially for protection of
applications against overcurrent. Therefore upper and lower limits for
current and voltage can be set which limit the range of the adjustable
set values.
For power and resistance only upper value limits can be set:
► How to congure the adjustment limits:
1. Tap the touch area in the main page.
2. Tap the white arrows , to select “3. Limits”.
3. In each case a pair of upper and lower limits for U/I or the upper limit for P/R are assigned to the rotary
knobs and can be set. Tap the selection area for another choice .
4. Accept the settings with .
The set values can be entered directly using the ten-key pad. This appears when the touch area
for rotary knob assignment is tapped (bottom center)
The adjustment limits are coupled to the set values. It means, that the upper limit may not be
set lower than the corresponding set value. Example: If you wish to set the limit for the power
set value P
SET
to 3000W and the currently adjusted value is 3500W, then the set value must
rst be reduced to 3000W or less. The same applies for minimum values .