Operator`s manual

TF-DSQ Probe Calibration and Deskew Fixture
TF-DSQ-OM-E RevC
17
Basic Probe Calibration
The TF-DSQ fixture is used to calibrate probes. Prior to beginning your measurements:
1. Assemble the TF-DSQ fixture
2. Attach the fixture, ideally to the Auxiliary Input or an unused oscilloscope channel
3. Access the Probe Calibration Menu
Now, follow these steps for each probe used:
1. Connect the probe to the oscilloscope channel
2. Attach the probe to the TF-DSQ fixture
3. Press the Full Calibration button in the Probe Calibration Menu.
4. Wait a few seconds as the probe is calibrated (calibration wizard closes at end of
calibration).
5. When the calibration completes, remove the probe from the TF-DSQ fixture
Now you are ready to probe the circuit and perform your measurements. If power is interrupted
during your measurements, reboot the oscilloscope and manually recall your settings.
Advanced Mode Probe Calibration Menu
The advanced mode is entered by checking the advanced mode box at bottom-right of the basic
probe calibration dialog:
Checking this box allows:
Calibration of gain/offset only
Calibration of deskew only
Access to the advanced menu (shown as a tab behind the Probes Cal dialog)
Gain/Offset Only Calibration
Pressing this button performs only the DC calibration of the probe on the specified channel. See
details of Probe DC Calibration.