User's Manual Part 3

Thermocouple Simulator:
7.3.3.3. Procedure
1. Connect the M.O.L.E.® Thermal Profiler to calibration standard.
2. Connect the M.O.L.E.® to a USB port with the USB computer interface cable.
3. Insert the M.O.L.E.® into the Thermal Isolation Box.
4. Start Hyperterminal.
5. Enter any Name for the Connection Description.