User's Manual Part 3

Allow the M.O.L.E.® Thermal Profiler to stabilize for 1/2 hour before calibration.
If after checking these possible sources of inaccuracy and the M.O.L.E.® Thermal Profiler
still needs to be calibrated, proceed as directed.
* IPTS-90 - International Practical Temperature Scale of 1990
7.3.3.1. Equipment Required
Equipment Required:
1. Voltage reference and an ice point reference.
Resolution: 1µV or better
Accuracy: 5µV or better
Output Imp: =10ohms
---------- OR ----------
Thermocouple Simulator
Resolution: 0.1°C
Accuracy: 0.25°C
Output Imp: =10ohms
2. Thermocouple harness (special limits of error)
3. Thermal Isolation Box (Thermal Barrier)
4. Thermocouple Simulator software program
Hyperterminal (Windows)
5. USB computer interface cable
7.3.3.2. Setup
The equipment you use for the calibration determines the setup procedure. The number
of thermocouples vary depending on the M.O.L.E.® Thermal Profiler. These examples
display the use of 6 channels.
Voltage reference and ice point reference: